Installation procedure

Mounting the shank on the probe head

Hold the shank in the recess on the top face of the MH8

Rotate the shank until the three screw holes are aligned

Fix the shank in place using only the M3 × 5 mm screws ﴾supplied﴿

Progressively tighten using the 2.5 mm A/F hexagon key (supplied)

Mounting the head on the CMM

The head will normally be attached to the quill of the CMM using a standard shank, specified by the CMM manufacturer.

Standard shanks are listed in the ‘Part number summary'.

Special or customised mounting arrangements are possible. Please contact your supplier or Renishaw, for details of our custom products

service.

The mounting must be absolutely rigid, as any movement during normal operation of the head will add errors to the measurement system.

If the head is replaced or repositioned in the quill, it will be necessary to requalify all of the head positions in current use.

Electrical connection

Connection is by means of a 5-pin DIN connector on the side of the head.

The pin numbering functions are given in the table below:

Pin number

Designation

1

LED cathode

2

Ground

3

LED anode

4

Probe circuit

5

Probe circuit

MH8 user's guide

www.renishaw.com

 

 

Issued 02 2020

8

Summary of Contents for A-1332-0002

Page 1: ...MH8 user s guide Documentation part number H 1000 5190 06 B MH8 user s guide www renishaw com Issued 02 2020 1 ...

Page 2: ...marks of their respective owners All trademarks and trade names are acknowledged WEEE The use of this symbol on Renishaw products and or accompanying documentation indicates that the product should not be mixed with the general household waste upon disposal It is the responsibility of the end user to dispose of this product at a designated collection point for waste electrical and electronic equip...

Page 3: ...r software without incurring any obligations to make changes to Renishaw equipment previously sold Packaging To aid end user recycling and disposal the materials used in the different components of the packaging are stated here Packaging component Material 94 62 EC code 94 62 EC number Outer box Corrugated fibreboard PAP 20 Packaging insert Corrugated fibreboard PAP 20 Bag Low density Polyethylene...

Page 4: ...t be used in a residential area or connected to a low voltage power supply network which supplies buildings used for residential purposes REACH regulation Information required by Article 33 1 of Regulation EC No 1907 2006 REACH relating to products containing substances of very high concern SVHCs is available at www renishaw com REACH China RoHS Contact Renishaw plc or visit www renishaw com China...

Page 5: ...ion of the stylus The head can carry probes fitted with a Renishaw M8 connector The TP20 and TP6 touch trigger probes are recommended The MH8 is available with a range of shanks to suit the mounting arrangement specified by the manufacturer of the CMM MH8 user s guide www renishaw com Issued 02 2020 5 ...

Page 6: ...tation is locked the probe is kinematically constrained in a highly repeatable spatial position This means that after initial qualification of the stylus tip in the positions required for gauging the workpiece the probe may be reorientated to any of the qualified positions without the need for requalification A red LED is provided on the front of the head to indicate the probe s status The LED may...

Page 7: ...TK4 MS5 shank A 1332 0007 MH8 TK4 MS7 shank A 1332 0008 MH8 TK4 MS9 shank A 1332 0009 MH8 TK4 MS10 shank A 1332 0010 MH8 TK4 MS13 shank A 1332 0011 MH8 TK4 MS15 shank A 1332 0014 MH8 TK4 MS17 shank A 1332 0012 MH8 TK4 D shank A 1042 1486 S1 C spanner A 1047 3932 S9 double ended C spanner M 5000 3540 S7 stylus tightening tool P TL03 0150 Hexagon key 1 5 mm AF TP2 TP6 adjustment MH8 user s guide www...

Page 8: ...ary Special or customised mounting arrangements are possible Please contact your supplier or Renishaw for details of our custom products service The mounting must be absolutely rigid as any movement during normal operation of the head will add errors to the measurement system If the head is replaced or repositioned in the quill it will be necessary to requalify all of the head positions in current...

Page 9: ...til it is finger tight Fit the S1 C spanner supplied to the location holes in the probe body and tighten by hand The recommended tightening torque is 0 3 Nm 0 5 Nm Refer also to the user s guide for the probe type being used Extension bars Probe reach may be extended by 50 mm using a PEL1 extension bar Use of the PEL2 PEL3 or PE series extension bars is not recommended MH8 user s guide www renisha...

Page 10: ...er to the fully clockwise position Qualify the stylus tip s according to the CMM supplier s instructions Change the orientation of the probe to the next desired position and qualify the stylus tip s Repeat the qualification process for all desired orientations and stylus tips Commence gauging ensuring that the correct qualification data is recalled for each head position Periodic requalification s...

Page 11: ...Dimensions MH8 user s guide www renishaw com Issued 02 2020 11 ...

Page 12: ... module and 10 mm stylus length 2 5 μm Technical data Range of articulation A axis 0 to 90 in 15 increments B axis 180 in 15 increments Maximum load TP6 no extension allowed TP20 or TP2 probe 50 mm extension PEL1 Operating temperature range 10 C to 40 C Storage temperature range 10 C to 70 C Probe connector Renishaw M8 1 25 mm 5 mm Electrical connector 5 pin DIN 180 Sealing IP30 Weight 205 g witho...

Page 13: ...common with all precision measuring equipment regular inspection and cleaning are recommended to ensure continued high performance Maintenance is limited to wiping the outer surfaces axes scale labels and mounting face of the probe connector with a clean dry cloth or proprietary cleaning material MH8 user s guide www renishaw com Issued 02 2020 13 ...

Page 14: ...the lock lever is turned fully clockwise Excessive force imparted to probe after locking Unlock and re lock the MH8 MH8 overloaded Use only with the specified probe and extension combinations Faulty probe Check by substitution MH8 worn or damaged Check by substitution or return to Renishaw or your supplier Axes rattle when rotated Possible causes Remedy MH8 not fully unlocked Turn lock lever fully...

Page 15: ...tails please visit our main website at www renishaw com contact Renishaw plc New Mills Wotton under Edge Gloucestershire GL12 8JR United Kingdom T 44 0 1453 524524 F 44 0 1453 524901 www renishaw com cmmsupport Issued 02 2020 ...

Reviews: