- 3 -
6. Cleaning /
Maintenance
The Bijou 90 requires no maintenance.
From time to time the burner should be cleaned.
Please observe the following cleaning instructions:
Never use water or compressed air to rinse
or blow out burners! This could flush out /
blow out the sealant grease causing the
burner to leak.
1. The maximum temperature for cleaning is 50 °C
(122 ºF).
2. Do not use cleaning agents because of the risk of
chemical reaction.
3. To clean the valve only use a duster or moist
leather (must not be dripping wet)!
4. To clean the burner close the shutoff valve, pull
the hose off the burner and wipe the burner clean
with a cloth or leather. The burner tubes can be
screwed off and cleaned separately.
5. Do not dismantle the burner!
6. Never drill out the burner jets.
7. Do not blow out the burner‘s inlet nozzle with
compressed air. This could blow away the sealing
grease and cause the burner to leak.
8. The burner jets must not be screwed out for
cleaning. The burner could start to leak and safe
use is no longer guaranteed.
If the burner nozzle becomes plugged,
disassemble the tubes than hold the burner
with the nozzle facing down. Use a hair dry-
er or other device generating hot air, melt
the wax plug and allow the molten wax to
completely run out.
If this method fails to clean the nozzle, it
must be either cleaned or replaced by an
authorized technician.
7. Switching to a
different type of gas
The Bijou 90 conforms with the regulations set out
in DIN 30665 Part 1 and enables a switch to another
type of gas by changing the set of burner nozzles.
Type of gas and power:
Type of gas
Rated power
Small
nozzle
Large
nozzle
Load
Load large
tube
Load small
tube
Product ref.
mark
Liquid gas
G 30 / 50 mbar
D 0.18 mm
D 0.34 mm
0.66 kW
0.5 kW
0.16 kW
960-0000
Natural gas E
G 20 / 20 mbar
D 0.32 mm
D 0.55 mm
0.66 kW
0.49 kW
0.16 kW
961-0000
• The natural gas E version is distinguished with the
sticker „Erdgas E 20 mbar“.
• The liquid gas version is distinguished with the
sticker „Flüssiggas 50 mbar“.
The predetermined lab burner can be switched over
to natural gas E and liquid gas by changing the set
of nozzles and its respective sticker. The relevant
conversion sets are given in the table above. Each
set consists of a small nozzle, a large nozzle and a
sticker naming the respective type of gas. The con-
version sets are featured in the spare parts list.
The changeover of the nozzles must only be
carried out by qualified specialists in ac
-
cordance with DVGW regulations (German
legislation - check local regulations).
EN
Summary of Contents for 9600000
Page 2: ......
Page 32: ...2 4 Bijou 90 4 1 4 2 4 3 Bijou 90 Bijou 90 90 5 o RU C B A...
Page 48: ...2 4 Bijou 90 4 1 4 2 4 3 Bijou 90 Bijou 90 90 5 UK C B A...
Page 52: ...2 4 Bijou 90 A B C A C 4 1 4 2 4 3 3 90 5 JA C B A...
Page 56: ...2 4 A B C C A 4 1 4 2 4 3 90 5 AR C B A...
Page 59: ......
Page 60: ......
Page 61: ......