EN
FAQ - English
For information on operating elements, see
For chapter indications, refer to the complete instruction manual on the internet.
Trouble
Possible Cause
Corrective Action
When the device is
switched on, there is a
loud vibration noise for
approx. 8 seconds.
• The suction unit performs an auto-
matic filter cleaning sequence.
•
This is functional and requires no inter
-
vention.
During an interruption in
operation, an acoustic
signal occurs 3 times and
then the filter cleaning
function is carried out.
•
After 16 hrs. constant operation (tur
-
bine running time) without having
been switched off in between, the
filter cleaning function is carried out.
•
Always switch appliances off at the on /
off switch at the end of the working day
(9,
).
When the unit is switched
on, the display dust drawer
an acoustic signal occurs
3 times.
•
The time period for emptying the dust
drawer has elapsed and the dust
drawer has not yet been emptied.
•
The emptying of the dust drawer has
not been confirmed.
•
Empty the dust drawer and confirm by
•
Empty the dust drawer and confirm by
lights up and an acoustic
signal occurs 3 times.
•
The automatic switch-on function
cannot be set. The difference be
-
tween the chosen rpm at the hand
-
piece and the stand-by current is too
low.
•
The automatic switch-on function
cannot be set.
•
Adjust the switch-on threshold to a higher
speed on the handpiece.
•
Use the unit in continuous operating
mode.
The error message
(26, Fig. 2) blinks.
•
The electronic device has become
too hot.
•
Switch the device off and allow it to cool
down.
•
Ensure sufficient cooling, e.g. with:
-
Chapter 4.2 installation instructions.
-
External ventilation (see chapter 4.6).
-
Change fine filter.
The suction performance is
insufficient.
• The set suction performance is too
low.
•
There is a blockage or leak in the
suction tube.
•
Dust drawer is not airtight.
•
The fine filter is full.
•
Select higher suction level.
•
Check suction tube.
•
Please observe the points in chapter 4.4
•
Check the dust drawer is in the correct
position (chapter 6.2).
•
Switch the unit off and back on in order to
initiate the filter cleaning function.
•
Change the fine filter (see chapter 6.3) (if
there is no improvement in suction perfor-
mance after the filter cleaning process).
Dust drawer is over full.
•
The time period for “empty dust draw
-
er” has been set too high.
•
Adjust to a lower time period (see chapter
6.2.1).
The signal to empty the
dust drawer shows even
though the dust drawer is
not yet full.
•
The time period for “empty dust draw
-
er” has been set too low.
•
Adjust to a higher time period (see chap
-
ter 6.2.1).
The electronic device con-
nected to the unit power
socket cannot be operated.
•
Suction is not switched on.
•
Switch the suction on.
- 1 -