- 16 -
6.2.4
Change Commpressed filter
⇒
Disconnect the appliance from the compressed air supply.
⇒
Roll away the O-ring seal.
⇒
Unscrew filter housing (hold up at the filter housing with the
screwdriver attached).
⇒
Using an Allan key (SW 5 mm) unscrew the sleeve (26).
⇒
Install a new piece of non-woven material (28) and sieve (27),
making sure to adhere to the correct.
⇒
Screw the sleeve back in and hand-tighten.
⇒
Screw filter housing together, replace the O-ring seal.
6.3 Self-Diagnosis
During self-diagnosis, various device functions are checked.
To carry out a self-diagnosis, see chapter 5.3.1.
This process takes approximately 30 seconds.
A blocked or defect suction hose can only be detected by the self-diagnosis and not when the
device is in action.
If an error is detected during the self-diagnosis, a fault code (see chapter 7.1) will be displayed.
Please continue as instructed.
6.4 Safety Mechanism
The protection of the suction device is achieved by two device
protection switches (10).
If an overload switch is released, it can be reset by pressing
the button back in.
Repeated triggering of a device protection switch is
caused by a defect in the suction device.
Send the device in for repair!
6.5 Factory Set Parameter
When the function “reset to factory set parameter” is activat-
ed, all individually set information is lost and overwritten by
factory set values.
⇒
Extraction off
⇒
and Press button (41, 43) and keep pressed.
⇒
Extraction on
⇒
and Press button (41, 43) and keep pressed
♦ An audible signal will confirm the factory set parameter is
saved and the display will indicate „rES“.
Factory set parameter:
Function / Feature
Adjustment Range
Factory Parameter
Extraction force, one suction
hose is open
1 - 3
2
Extraction force, both suction
hoses are open
1 - 3
2
Sensitivity of dust bag fill level
Recognition
1 - 9
5
Audible Signal
On / Off
On
Automatic start-up threshold
1 - 100 W
8 W
Shut-off delay A or B,
one suction hose is open
3 - 30 Sec.
3 Sec.
Shut-off delay A / B,
both Suction hoses are open
3 - 30 Sec.
15 Sec.
Only for devices with a software version up to 1.00.15 (see chapter. 5.3.1):
Diagnosis - mode
(after switching on)
On / Off
Off
Fig. 22
26
27
28
Fig. 23
10
Fig. 24
43
41
EN