ENGLISH

1. Introduction

We are pleased with your decision to purchase the 

Basic mobil

. The mobile sandblasting unit is connect-

ed to the compressed air or to the recyclable blaster 

Vario-Jet

 risp. 

Vario basic 

.

2. Application Area

The 

Basic

 line of sandblasting units are designed 

for use in dental laboratories to remove investment 

residues, oxides on cast objects, and for surface 

treatment.

3. Safety Information

•  Always disconnect the unit from the com-

pressed air and power sources prior to begin-

ning any maintenance work.

•  When working on the tank (filling up, clean

-

ing, maintenance) wear protective glasses for 

protecting your eyes.

•  Any residue of abrasive material on the gas

-

ket may lead to leakage and early wear of the 

gasket. After filling up clean the thread and the 

gasket and close the lid properly.

•  Solvents and tensides can create micro-crack

-

ing in the plastic (danger of explosion!). Clean 

the tank and the lid only with a dry cloth. Do 

not write or stick something on the tanks.

•  Check the tanks and the lids regularly on dam

-

ages and replace them in case of doubt.

•  Check the tank cover for correct seating prior 

to beginning operation. Covers that are not se

-

curely closed can suddenly be blown off as the 

tank pressure increases. The resulting flying 

parts and abrasive represent a serious hazard.

• 

Never

 operate the blasting units without suit-

able dust extraction and appropriate protective 

gear, as this could result in health hazards

. The 

type of dust extraction should be selected on 

the basis of the dust being generated. Follow 

here absolutely EN 60335-2-69 appendix AA or 

ask the appropriate authorities.

•  Improper use can result in the risk of eye or 

skin injuries.

•  Never direct the blasting material towards eyes 

or unprotected areas of skin!

  Caution: Failure to wear proper eye 

protection can result in eye injury due 

to particles suspended in the air. Always 

wear proper eye protection to protect 

your vision!

•  Do not employ the foot switch when only the 

compressed air supply hose is connected. The 

loose hose could wrap around itself and cause 

serious injury.

•  Only spare parts and accessories supplied or 

authorized by Renfert GmbH may be used with 

this product. If other spare parts or accessories 

are used, this could have a detrimental effect 

on the safety of the device, increase the risk of 

serious injury and lead to damage to the envi

-

ronment or the device itself.

•  It is the responsibility of the operator that na-

tional regulations during operation and regard-

ing a repeated safety inspection of electrical 

equipment are complied with.  

For Germany these are BGV A3 in relation with 

VDE 0701-0702.

3.1 Liability Exclusion

Renfert GmbH shall be absolved from all claims 

for damages or warranty if:

•  The product is employed for any purposes 

other than those cited in the operating instruc-

tions;

•  The product is altered in any way other than 

those alterations described in the operating 

instructions;

•  The product is repaired by other than an autho-

rized facility or if any but Renfert OEM parts are 

employed;

•  The product continues to be employed, despite 

obvious safety faults;

•  The product is subjected to mechanical jarring 

or is dropped.

Summary of Contents for 2914

Page 1: ...gsanleitung Instruction manual Mode d emploi Istruzioni d uso Instrucciones de servicio Инструкция по эксплуатации 取扱説明書 사용 설명서 Made in Germany Basic mobil Nr 2914 219093 03032016 B Ideas for dental technology ...

Page 2: ...1 2a 2c 4 6 5 3 2b ...

Page 3: ...7 9a 10 12 13 11 9b 8 ...

Page 4: ...14 ...

Page 5: ...raction should be selected on the basis of the dust being generated Follow here absolutely EN 60335 2 69 appendix AA or ask the appropriate authorities Improper use can result in the risk of eye or skin injuries Never direct the blasting material towards eyes or unprotected areas of skin Caution Failure to wear proper eye protection can result in eye injury due to particles suspended in the air Al...

Page 6: ...The sandblasting unit is now operational 5 Cleaning Maintenance Use only solvent free cleansers e g soapy water Wipe out the tanks and lids only with a dry cloth 5 1 Condensation 1 Inspect the water separator 2 Close the shut off valve Fig 11 3 Drain any water off through the valve Fig 12 employing the foot switch Fig 6 5 2 Blasting Nozzle Replacement 1 Refer to insert no 21 9162 for installation ...

Page 7: ...acity 1000 ml 34 fl oz Weight empty approx 2 5 kg 5 5 lbs 9 Notes Flat blasting with sheet dies Art No 90002 1207 0 3 x 3 5 mm 0 012 x 0 138 inch Art No 90003 1739 1 5 x 3 5 mm 0 059 x 0 138 inch When blasting glass ceramics refer to manufac turer s notes 10 Standard Delivery 1 Basic mobil 1 Foot switch only with no 2914 3050 2914 3250 1 Connection set 1 Set of Operating Instructions with attachme...

Page 8: ... or moist sand Mixing chamber filter combination blocked Abrasive contains impurities or is moist Cleaning the dosing nozzle see chapter 5 4 Change filter cartridge Wrench SW14 Replace the abrasive Unit blasts continu ously Foot switch connected incorrectly Valve in foot switch is defective Inspect correct the foot switch connection Replace foot switch Compressed air leaking from the foot switch P...

Page 9: ...ert USA 3718 Illinois Avenue St Charles IL 60174 USA Tel 1 6307 62 18 03 Fax 1 6307 62 97 87 www renfert com richardj renfertusa com USA Free call 800 336 7422 Ideas for dental technology Hochaktuell und ausführlich auf Up to date and in detail at Actualisé et détaillé sous Aggiornato e dettagliato su La máxima actualidad y detalle en Актуально и подробно на www renfert com ...