OB1203 Certification Preparation, Setup, and Test Procedure
X0119984 Rev.1.0
Mar 11, 2021
Page 1
© 2021 Renesas Electronics
Overview
SpO2 measurements are affected by two variations in the human body. The first is skin color. The second is the
blood perfusion, which causes optical modulation and varies based on the person’s temperature and health.
Changes in these variants affect SpO2 measurements.
In this pre-calibration test procedure, both of these variables are simulated and the oximeter output is monitored.
Renesas provides test stimuli simulating various conditions that represent the majority of the human population.
Therefore, a successful execution in the actual human subject testing certification testing can be assumed.