OB1203 Certification Preparation, Setup, and Test Procedure
X0119984 Rev.1.0
Mar 11, 2021
Page 16
7.3
SP02 Saturation Value Test before Calibration
This procedure tests whether the DUT is properly reporting values from 100% to 70%. This will establish the
correction values (R-curve) for the DUT on the simulated test. It does not check the actual accuracy on human
test subjects.
Set the AECG100:
a. Set the default calibration curve of the OB1203.
SpO2 = 107.36 - 10.29 * R - 28.15 * R
2
(default calibration curve)
b. Set the AECG100 SpO2 settings to:
o
Red DC of 2000mV
o
IR DC of 1000mV
o
Click both Lock DC boxes
o
Red AC of 15mV
o
IR AC of 18.7mV
c. Click the radio button for PI(R).