REITEL STEAMY PR Operating Instructions Manual Download Page 7

 

 7 

  Safety Notes 

  General Safety Notes 

  All equipment undergoes final inspection and testing before leaving our factory. However, 

if equipment is damaged during shipping or through any other circumstances, it must not 

be used in service. Please check the device regularly for damages. Damaged parts 

should be replaced immediately. 

  All power/air/water/gas sources should be rated for the demands of the device as listed in 

the technical data sheets. Maximum ratings must not be exceeded. All VDE (Verein 

Deutscher Elektro-Ingenieure) standards should be followed. Equipment should be 

grounded according to standards. 

  Adherence to valid regulations regarding necessary checks (e.g. according to Accident 

Prevention & Insurance or electrical equipment) is exclusively taken over by the 

buyer/user of this product. 

  This device has been designated to be used only for its intended purpose. All other 

applications must be prohibited for safety reasons as otherwise the manufacturer’s 

warranty expires. 

  The device should be set-up in a clean, well-ventilated room, if necessary under a hood. 

  Supporting surfaces should be straight, flat and adequately stable. 

 

Don’t place anything on top of the machine. 

  To avoid injury, appropriate protection should be worn (e.g. gloves, safety glasses, safety 

shoes, hearing protection etc.) 

  If the machine is defective or working improperly it must be taken out of service. In case 

of further use of the machine the owner will be held responsible for any consequential 

damage. 

  To prevent accidents from water, fire or explosion, all power/air/water/gas sources should 

be turned off after working hours and during breaks, and secured against unauthorized 

access. 

  When carrying out maintenance works or repairs or when opening the device always turn 

off the main power switch and unplug the whole unit. 

  Repairs should be carried out only by the manufacturer or by qualified and certified 

personnel appointed according to chapter 7 

– warranty. 

  In case of leaking water out of housing or tank shut down device and call service. 

  Check all hose joints for tightness, if necessary fasten clamps. 

 

 

Summary of Contents for STEAMY PR

Page 1: ...erating Instructions LABORATORY EQUIPMENT IN STAINLESS STEEL STEAMY PR REITEL Feinwerktechnik GmbH Senfdamm 20 49152 Bad Essen Phone 49 0 5472 9432 0 Fax 49 0 5472 9432 40 www reitel com info reitel c...

Page 2: ...the device please check carefully whether transportation damages of any kind have occurred If that is the case the carrier or manufacturer must be notified within 3 days The company will not be respon...

Page 3: ...TION 6 VERSION WITH FLAT NOZZLE 6 TECHNICAL DATA 6 SAFETY NOTES 7 GENERAL SAFETY NOTES 7 SPECIFIC SAFETY NOTES 8 INSTALLATION 9 OPERATION 9 CHRONOLOGICAL RUN OF OPERATION 9 DECALCIFYING THE DEVICE WHE...

Page 4: ...Order No 10205000 Water softening device DELIMER COMFORT Operating Elements 1 Steam valve 2 POWER switch 3 Swith to activate hose heating HEATING 4 Indicator lamp low water level 5 Heating control 6 T...

Page 5: ...5 Device view Backside 1 13 11 5 4 6 7 3 9 10 2 12 14 8...

Page 6: ...lat nozzle Technical Data Line voltage 230 V 50 60 Hz Current consumption 2 300 W Height width depth 420 300 350 mm Weight 14 kg Pressure 4 bar Temperature 150 C Volume 5 l Steam level 1 Filling manua...

Page 7: ...e the manufacturer s warranty expires The device should be set up in a clean well ventilated room if necessary under a hood Supporting surfaces should be straight flat and adequately stable Don t plac...

Page 8: ...cale deposits is not covered by warranty Decalcification is depending on water hardness every one or two weeks Verwenden Sie hierf r handels bliche Entkalkungsmittel und beachten Sie die Herstellerang...

Page 9: ...the device on with the switch POWER 2 Set heating control 5 to max If you don t set heating controller to max the low water level indicator lamp lights up 4 despite a filled tank However the device is...

Page 10: ...vice cool itself down and empty the tank 3 Fill the tank with decalcifier according to manufacturer s instructions 4 Close tank cap 6 5 Turn switch POWER 2 on As soon as the pressure starts building u...

Page 11: ...or seal No steam a Cold water was refilled and device is heating up b Device is calcified blocked valve switches or valves defective Call customer services No steam pressure 1 Regulator 5 was not set...

Page 12: ...g the device when not using distilled water Clean the housing regularly with a moist cloth Note Clean and protect your device regularly with the REITEL stainless steel care spray Order No 90405000 to...

Page 13: ...ssed to the distributor the product was purchased from or directly to our service department Contact REITEL Feinwerktechnik GmbH Senfdamm 20 49152 Bad Essen Germany Phone 49 0 5472 9432 32 Service Dep...

Page 14: ...ages due to faulty installation or insufficient maintenance of the product e g decalcification Damages due to natural disaster e g lightning water fire act of nature beyond control war inadequate line...

Page 15: ...dvised to backup essential programs and data before sending for service 12 This warranty shall be interpreted in accordance with German law to the exclusion of UN sales law Place of jurisdiction is ou...

Page 16: ...16 Disposal Regulations According to valid laws our devices are disposed of according to B2B Regulations for proper disposal Please contact your local dealer for additional information...

Page 17: ...ng Rubber ring 1 St pc E6227700 Heizk rper inclusive Flansch und Dichtung Heating element incl flange and seal 1 St pc E7327600 Isoliermatte nicht gezeichnet Isolation mat not drawn 1 St pc E6021200 K...

Page 18: ...r Schlauchheizung 24 V 75 W Transformer for hose heating 24 V 75 W 1 St pc E6175500 Verschraubung UNI VS 13 ohne DK 030 Screw joint UNI VS 13 without DK 030 1 St pc E7090000 Verschraubungen bitte For...

Page 19: ...19 EU DIRECTIVE...

Page 20: ...pliance with Quality Management ISO 13485 2016 incorporating established standards REITEL Feinwerktechnik GmbH Senfdamm 20 49152 Bad Essen Phone 49 0 5472 9432 0 Fax 49 0 5472 9432 40 www reitel com i...