IMPORTANT SAFETY INSTRUCTIONS 

 

 

 

 

 

 

 

 

 

 

 

 If the arm board is to be used with diathermy procedures, the pad must be fixed to the arm 

board plate by f.ex. Velcro straps to prevent it from sliding. The whole arm board including 
the maneuver lever must be covered by a sterile bag or anything else to prevent the patient 
from getting in contact with metal parts. 

 

 

The personnel, responsible for adjustment of the arm support, must be sure of there are no 
human parts in the risk area when the arm board is adjusted.  

 

 

   

 

 

 

 

 

 

   

 

 

 

The arm support is designed for a Safe Working Load (SWL) of 15 kg evenly applied. To 
keep the safety factor according to the European Standards EN 60601-1 and EN 60601-2-46 
the Safe Working Load shall not be exceeded. Overload can cause injuries to the patient, or 
damage the equipment.15 kg SWL is corresponding to a patient weight of 200 kg. 

 

 

 

Before every use with a patient, a function control of the arm support must always be done. 
 

 

Service and preventing maintenance shall be done by authorized personnel and according to 
issued intervals. 
 
 

 

 

 

 
 

RISK AREA 

Maximum load 15 kg evenly applied 

Summary of Contents for 10-380-KL

Page 1: ...10 380 KL Edit 3 Page 1 5 USERS MANUAL ARM SUPPORT 10 380 KL GMDN 10184 PID 122493 ...

Page 2: ... manual close to the equipment INSTALLATION At the unpacking check for any kind of damage on the equipment Also check that all parts according to the delivery note are received The packing material consists of cardboard and plastic which all can be recycled GENERAL DESCRIPTION To adjust the top put the hand at the end of the top grasp the T shaped lever with the fingers and pull the lever upwards ...

Page 3: ...ere are no human parts in the risk area when the arm board is adjusted The arm support is designed for a Safe Working Load SWL of 15 kg evenly applied To keep the safety factor according to the European Standards EN 60601 1 and EN 60601 2 46 the Safe Working Load shall not be exceeded Overload can cause injuries to the patient or damage the equipment 15 kg SWL is corresponding to a patient weight ...

Page 4: ...he arm support shall be carefully cleaned moving parts shall be lubricated the locking function shall be tested and if necessary the ball joint shall be adjusted BALL JOINT ADJUSTMENT Adjusting of the release point Loosen the set screw adjust the release lever by tighten or loosen the release lever adjustment screw After correct adjusting tighten the set screw Adjusting the locking force For highe...

Page 5: ... 1kg Safe Working Load 15 kg Max patient weight 200 kg Box 240 Eriksbergsvägen 32 A 734 25 Hallstahammar Sweden Phone 46 220 43399 Fax 46 220 29016 E mail info reison se www reison se 1 Ball joint device 11914 1pc 2 Top plate 10514 1pc 3 Screw MFX Z A2 POZ M4x10 4pcs 4 Plastic plug IKP 2015 17 S flanged 1pc ...

Reviews: