IMPORTANT SAFETY INSTRUCTIONS
❖
If the arm board is to be used with diathermy procedures, the pad must be fixed to the arm
board plate by f.ex. Velcro straps to prevent it from sliding. The whole arm board including
the maneuver lever must be covered by a sterile bag or anything else to prevent the patient
from getting in contact with metal parts.
❖
The personnel, responsible for adjustment of the arm support, must be sure of there are no
human parts in the risk area when the arm board is adjusted.
❖
The arm support is designed for a Safe Working Load (SWL) of 15 kg evenly applied. To
keep the safety factor according to the European Standards EN 60601-1 and EN 60601-2-46
the Safe Working Load shall not be exceeded. Overload can cause injuries to the patient, or
damage the equipment.15 kg SWL is corresponding to a patient weight of 200 kg.
❖
Before every use with a patient, a function control of the arm support must always be done.
❖
Service and preventing maintenance shall be done by authorized personnel and according to
issued intervals.
RISK AREA
Maximum load 15 kg evenly applied