Cleaning
External Cleaning
WARNING: DO NOT IMMERSE THE CONNECTOR OR THE MAIN UNIT IN ANY LIQUID OR DAMAGE TO THE UNIT WILL
OCCUR.
Clean the external surfaces of this instrument using a clean, soft cloth moistened with a mild detergent solution (1 cc of
liquid dish soap to one liter of clean, filtered water (filtered below 5 microns)).
Probe Cleaning
WARNING: IN ORDER TO PREVENT PATIENT-TO-PATIENT TRANSFER OF INFECTION, AFTER EACH USE DISINFECT THE
PROBE FOLLOWING ACCEPTED LOCAL CLINICAL PROCEDURES REGARDING THE USE OF
DISINFECTANTS. ANY CLINICALLY APPROVED CHEMICAL DISINFECTANT CAN BE USED.
WARNING:
DO NOT ATTEMPT TO USE THE IOPAC IF THERE IS ANY INDICATION THE PROBE TIP, BODY, OR CABLE
HAS BEEN DAMAGED AND/OR THEIR PHYSICAL INTEGRITY HAS BEEN COMPROMISED OR INJURY TO THE
CORNEA MAY OCCUR.
CAUTION:
DO NOT USE THE PROBE AS A STYLUS OR DAMAGE TO THE PROBE MAY OCCUR.
CAUTION:
DO NOT STERILIZE THE PROBE USING ETO, AUTOCLAVING, OR IRRADIATION OR DAMAGE TO THE PROBE
WILL OCCUR.
CAUTION:
IF THE PROBE TIP HAS MADE CONTACT WITH ANYTHING BETWEEN APPLANATIONS, REPEAT THE
CLEANING PROCEDURE.
Clean the probe using the following steps:
1.
Wipe the complete probe with a cotton swab soaked in 70% isopropyl alcohol.
2.
Immerse the probe tip for 10 minutes in 70% isopropyl alcohol or an equivalent locally approved disinfectant.
3.
Rinse the end of the probe thoroughly with sterile distilled water.
4.
Dry the probe with a clean, lint-free, cloth or tissue.
5.
Store the probe in its provided holder on the unit when it is not in use so that it is safely stored.