Perform the following steps to take a measurement of the patient’s eye.

  Set the height of the table so the canthus 

marks on the sides of the instrument are 

level with the patient’s eyes. The patient 

should be seated comfortably and posi-

tioned in a way that enables them to lean 

forward without straining.

  Slide the headrest fully to the left or right.

  Have the patient locate the red LEDs 

surrounding the air tube and lean forward 

so that the center of their forehead rests 

firmly in the middle of the rubber forehead 

pad.   

  Make sure the patient is able to see all, 

or some, of the red lights.  Explain to the 

patient that a green light will move into view 

during measurement.  Ask the patient to 

blink a few times then hold both eyes open 

when the green light appears.  

  Touch the MEASURE  or TRIPLE MEA-

SURE icon to begin the measurement 

process.

Note:

 If the instrument is unable to complete its alignment and measurement process, it 

may be necessary to:

• Reposition the patient and obtain optimal alignment

• Ask the patient to remain still and try not to move

• Remind the patient to keep both eyes wide open and to move his/her chin as close to 

the instrument as possible.

Reichert 7CR Quick Reference Guide

Patient Positioning & Alignment Guide

Proper

 patient alignment

(chin close to unit)

Improper

 patient alignment

(chin moved away from unit)

Illuminated

nosepiece

Air tube

Illuminated Nosepiece

Slide headrest left or right

Canthus Marks

Slide Headrest