Mary Lift User Manual
03-2015
TD 5.6-001 PR 3500 200 E - ENG ver.4
Side 16 af 20
Annex C Recommended annual service inspection
(Complete/adapt as appropriate)
Date:
Product ID / serial no.
Carried out by:
Service undertaken by:
Nr.
Check
Remarks
Visual control of product
1
-
wear
2
-
distortion
3
-
general damage
4
-
rust
5
-
stability – no loose parts
6
-
wheels run easily
7
-
wheel brakes function
8
-
wear of lift hooks
9
-
connection between hook and sling
Actuator
10
-
test top stop position
11
-
Actuator needs replacing after 11,000 lifts or at least
every 5 years
Control all functions in normal use (one full cycle with
maximum load, no problems or abnormal sounds)
12
-
Control emergency down
13
-
Control emergency stop
14
-
Control battery charge indicator
As a standard, the control box should be serviced every 12
months/8000 lifts, whichever comes first.