Function
Range & Resolution
Accuracy (% of reading)
DC.Current
40.00.ADC
0~20.00ADC
±.(2.5%.+.6.digits)
20.00~40.00ADC
±.(3%.+.6.digits)
400.0.ADC
0~300.0ADC
±.(2.5%.+.6.digits)
300.0~400.0ADC
±.(3.5%.+.6.digits)
AC.Current
40.00.AAC
0~20.00AAC
±.(3%.+.10.digits)
20.00~40.00AAC
±.(5%.+.10.digits)
400.0.AAC
0~300.0AAC
±.(3%.+.10.digits)
300.0~400.0AAC
±.(5%.+.10.digits)
DC.Voltage
4.000.VDC
±.(0.8%.+.3.digits)
40.00.VDC
±.(1.5%.+.3.digits)
400.0.VDC
600.VDC
±.(2.0%.+.3.digits)
AC.Voltage
400.0.mVAC
±.(1.0%.+.10.digits)
4.000.VAC
±.(2.0%.+.5.digits)
40.00.VAC
400.0.VAC
600.VAC
±.(2.0%.+.5.digits)
Resistance
400.0 Ω
±.(1.0%.+.4.digits)
4.000KΩ
±.(1.5%.+.2.digits)
40.00KΩ
400.0KΩ
4.000MΩ
±.(2.5%.+.3.digits)
40.00MΩ
±.(3.5%.+.5.digits)
Capacitance
40.00nF
±(5.0%.r.30.digits)
400.0nF
±(3.0%.r.5.digits)
4.000μF
±(3.5%.r.5.digits)
40.00μF
100.0μF
±(5.0%.r.5.digits)
Frequency
5.000Hz
±(1.5%.r.5.digits)
50.00Hz
±(1.2%.r.2.digits).
Sensitivity:.5~5kHz:10Vrms.min..
5kHz~150kHz:[email protected]%.
to.80%.duty.cycle
500.0Hz
5.000kHz
50.00kHz
150.0kHz
Duty.Cycle
0.5.to.99.0%
±(1.2%.r.2.digits)
Pulse.width:.100μs.-.100ms,.Frequency:.5Hz.to.150kHz;.
Sensitivity:.5~5kHz:10Vrms.min.5kHz~150kHz:40Vrms.min.