5
Pantone 534 Blue
Pantone 123 Yellow
Pantone 485 Red
Pantone 123 Yellow
Pantone 534 Blue
Black
Rich Black -
20/20/20/100
Blue - 100/80/30/5
Yellow - 0/27/100/0
Red - 10/100/100/5
Yellow - 0/27/100/0
Blue - 100/80/30/5
Only if you REALLY need them:
Pantone 534 Blue - 100/80/30/5
Pantone 485 Red - 10/100/100/5
Pantone 123 Yellow - 0/27/100/
INSTRUMENTS
INSTRUMENTS
Only if you REALLY need them:
• To calibrate the scale start with the scale off and poceed to turn it on by
pressing the
ON / OFF
button.
• When the digital screen displays 0, press and hold the
UNITS
button for
5 seconds, until CAL is displayed, release the
UNITS
button.
• Press the
UNITS
button again, the display will flash a weight. Add
the calibration (2.5 kg) weight on the middle of the platform. The LCD
display reads [
PASS
] and will return to the normal weighing mode.
• The scale is calibrated and can be turned off by pessing the
ON / OFF
button to turn off the scale.
NOTE: Repeat this process if necessary.
Troubleshooting
The primary reasons for inaccuracy or malfunction is low batteries,
incorrect calibration, overload or operating on an unstable surface.
Please keep these in mind and maintain and operate your scale properly.
The scale is a precise instrument and must be handled with the upmost
care and caution.
DISPLAY CODE
CODE DESCRIPTION
OUT2
Keep scale clean under platform then
calibrate and reboot the scale
UNSE
Put scale in more stable position
EEEEE
Scale is overloaded and will have to
be recalibrated
LO or NO SIGNAL
Batteries to be recharged
ERROR2
Scale is overloaded and will have to be
recalibrated. Scale may be damaged, check
the sensor and electrical
AU70
One or more buttons may be stuck
For service on this or any other REED product or information on other
REED products, contact REED Instruments at [email protected].