Measuring DC voltage

1.  Insert one end of the test lead into the 

input (INPUT) terminal of the meter and 
connect the other end to the output 
terminal of the instrument under test,  
as shown in the diagram to the right. 

2.  Press INPUT/OUTPUT button to select 

the Input function. ‘INPUT’, ‘ON’, ‘mV’ 
will be displayed.

3.  Starts the measurement process by  

displaying ‘000.00’ which indicates  
waiting and then displays measurement result. (Refresh rate is twice 
a second, and if the measured value exceeds the measurement 
range, the LCD displays ‘OL’.)

Measuring thermocouples

1.  Insert one end of the test lead into the input (INPUT) terminal of the 

meter and connect the other end to the output terminal of the instru-
ment under test, as shown in the diagram above. 

2.  Press INPUT/OUTPUT button to select the Input function. 
3.  Press the FUNC (function) button to select Thermocouple function.  

‘R’ and ‘°C’ will appear on the display..

4.  Press the RANG (range) button to select the type of thermocouple.
5.  Starts the measurement process by displaying ‘000.00’ which indi-

cates waiting and then displays measurement result. (Refresh rate is 
twice a second, and if the measured value exceeds the measurement 
range, the LCD displays ‘OL’.)

6.  Press the °C/°F button to select °C or °F unit of measure.

Battery Replacement

Note: 

The meter is powered by two AA alkaline batteries (LR6).

Note: 

Ensure that the battery’s positive and negative terminals mirror the 

illustration in the battery compartment when replacing them.

Note: 

New and old batteries cannot be mixed. 

continued ...

 

1.888.475.5235

[email protected]

REED-Direct.com

Summary of Contents for R2810

Page 1: ...ments www com Model Instruction Manual Black Pantone 534 Blue 100 80 30 5 Pantone 485 Red 10 100 100 5 Pantone 123 Yellow 0 27 100 Thermocouple Calibrator R2810 1 888 475 5235 info REED Direct com REED Direct com ...

Page 2: ...m Pantone 534 Blue 100 80 30 5 Pantone 485 Red 10 100 100 5 Table of Contents Safety Information 3 Features 3 Specifications 4 5 Instrument Description 6 7 Operating Instructions 8 11 Power ON OFF 8 Auto Power Off 8 Output Function 8 10 Simulating output DC voltage 9 Simulating output from thermocouple 9 10 Measure Function 10 11 Measuring DC voltage 11 Measuring thermocouple 11 Battery Replacemen...

Page 3: ...terminals or between any terminal and ground terminal Caution The meter case is not to be opened unless by a qualified technician Use a damp cloth with neutral detergent for cleaning the meter period ically Do not use abrasives or solvents Note For optimal accuracy allow the instrument to warm up 5 minutes before operating If the automatic reference junction temperature compensation of the instrum...

Page 4: ...0 to 1760 C 1 C 0 05 rdg 3 C 100 C 0 05 rdg 2 C 100 C S 20 to 1760 C 1 C B 400 to 1800 C 1 C 0 05 rdg 3 C 600 C 0 05 rdg 2 C 600 C E 200 0 to 1000 0 C 0 1 C 0 05 rdg 2 C 100 C 0 05 rdg 1 C 100 C K 200 0 to 1370 0 C 0 1 C J 200 0 to 1200 0 C 0 1 C T 200 0 to 400 0 C 0 1 C N 200 0 to 1300 0 C 0 1 C Measure Function Output Range Output Range Resolution Accuracy DCV 100mV 10 00mV to 110 00mV 0 01mV 0 ...

Page 5: ...en any two terminals or between any terminal and ground Temperature Coefficient 0 1 dedicated accuracy C 5 to 18 C 28 to 40 C Operating Temperature 0 C to 50 C Operating Humidity 80 RH Storage Temperature 10 C to 55 C Storage Humidity 90 RH Power Supply Two 1 5V alkaline batteries LR6 Dimensions 180 L 90 W 47 D mm with holster Weight 500g with holster Includes Set of industrial test leads alligato...

Page 6: ...antone 534 Blue 100 80 30 5 Pantone 485 Red 10 100 100 5 Instrument Description 1 LCD screen 2 ON OFF button 3 Input Output button 4 Zero Reset button 5 Function Selection Output button 6 Range Selection button 7 RJ ON button T C cold junction compensation button 8 Output Value Setting button 9 Output Digit Selection button 10 Input Output terminal continued 5 6 8 9 10 1 2 3 4 7 continued continue...

Page 7: ...ndicates that the instrument is in an input state Indicates that the instrument is in a calibration state Indicates that the instrument is in a calibration state denoting that the zero point or the full scale point is now in calibration Indicates that the instrument is performing its reference cold junction compensation Indicates that the battery power is low and needs to be replaced Indicates cur...

Page 8: ...the instrument will automatically turn off after 15 minutes of inactivity To turn off this feature 1 Turn the instrument off 2 Press the ON OFF button to display the full screen 3 Press the RANG button when the instrument is in the maintenance state AP XX will appear on the display 4 Press the button when AP ON is displayed AP OF will now appear indicating that the automatic power off is turned of...

Page 9: ... the value will keep varying 6 Press the ZERO button and the output will be set to 00 00mV or 0 0000V Simulating thermocouple TC output 1 Insert one end of the test lead into the OUTPUT terminal of the calibrator and connect the other end to the input terminal of the instrument under test See diagram above 2 Press the INPUT OUTPUT button to select the Output function 3 Press the FUNC function butt...

Page 10: ...e is a change in the operating ambient temperature do not start the operation until the built in compensating sensor has become stable about 10 minutes If there is no need for the calibrator to perform the function of automatic reference junction compensation press the RJ ON button and the symbol RJ ON will no longer appear in the display Measure Function Warning The maximum voltage allowed betwee...

Page 11: ...ays OL Measuring thermocouples 1 Insert one end of the test lead into the input INPUT terminal of the meter and connect the other end to the output terminal of the instru ment under test as shown in the diagram above 2 Press INPUT OUTPUT button to select the Input function 3 Press the FUNC function button to select Thermocouple function R and C will appear on the display 4 Press the RANG range but...

Page 12: ...eveal the battery compartment door which can be removed using a Phillips head screwdriver 3 Replace the batteries in the lid of the battery compartment 4 Reinstall the compartment lid by ensuring that battery terminals touch the unit s contact points and snap into place Note Ensure battery door is closed and latched before using the meter To ensure proper operation please wait 5 seconds before tur...

Reviews: