REED R2800 Instruction Manual Download Page 9

9

Auto Power-Off

As a default the instrument will automatically turn off after 15 minutes of 
inactivity. To turn off this feature:
1.  Turn the meter off.
2.  Press the POWER button (to display the full screen).
3.  Quickly press the 

RANG

 button when the instrument is in the 

maintenance state as indicated by "1.8.8.888".

4.  "AP-XX" will now appear on the display.
5.  Press the   button to toggle between "AP-ON" & "AP-OF". "AP-OF" 

indicates that the no automatic power-off function is disabled, while 
"AP-ON" indicates the automatic power-off function is enabled.

6.  Press the 

ZERO

 button to save the required setting.

7.  Press and hold the POWER button to exit the maintenance state and 

turn the instrument off.

Output Function

•  The output terminal of the instrument can produce DC voltages set by 

the user or can simulate resistance.

Do not apply any voltage to the output terminal during the operation. 

If any improper voltage is applied to the output terminal, it will cause 

damage to the internal circuit.

Simulating Output from DC Voltage

1.  Insert one end of the test lead into 

the output (TC/mV) jack of the 
meter and connect the other end 
to the input of the instrument under 
test, as shown in the diagram to 
the right.

2.  Press the 

FUNC

 button to select 

"DC Voltage Output". "V" will 
appear on the display.

continued...

 

REED Instruments

1-877-849-2127 | [email protected] | www.reedinstruments.com

Summary of Contents for R2800

Page 1: ...Temperature Simulator Instruction Manual R2800 REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com ...

Page 2: ...r ON OFF 8 Auto Power Off 9 Output Function 9 Simulating Output from DC Voltage 9 10 Simulating Output from Thermocouple TC 10 11 Simulating output from thermal resistance or RTD 11 12 Battery Replacement 12 Applications 13 Accessories and Replacement Parts 13 Product Care 13 Product Warranty 13 Product Disposal and Recycling 14 Product Support 14 REED Instruments 1 877 849 2127 info reedinstrumen...

Page 3: ...oduct other than for the purpose of replacing batteries may cause damage that will not be covered under the manufacturer s warranty Servicing should only be provided by an authorized service center To avoid injury to the user or damage to the instrument please read the safety information below before initial use Do not operate the instrument around flammable or explosive gas vapor or dust Never ap...

Page 4: ...y indicator and auto shut off Included Test Leads Alligator Clips Thermocouple Adapter Protective Holster Batteries Specifications Output Functions Accuracy specified at 23 C 5 C 75 RH for a period of one year after calibration Output Range Output Range Resolution Accuracy DCV 100mV 10 00 to 110 00mV 0 01mV 0 05 rdg 30mV 1000mV 100 00 to 1100 00mV 0 1mV 0 05 rdg 0 3mV OHM 400Ω 0 0 to 400 0Ω 0 1Ω 0...

Page 5: ...to 150 C 0 1 C Pt10 385 200 0 to 850 C 0 1 C 0 05 rdg 0 6 C Pt100 385 200 0 to 850 C 0 1 C Pt200 385 200 0 to 630 C 0 1 C 0 05 rdg 0 6 C Pt500 385 200 0 to 630 C 0 1 C Pt1000 385 200 0 to 630 C 0 1 C Notes Excludes accessory lead resistance Range of excitation current 0 5mA to 3mA Max output voltage 2V The accuracies indicated do not include the error of internal temperature compensation caused by...

Page 6: ...es after 15 minutes off Low Battery Indicator Yes Replaceable Test Leads Yes Product Certifications CE Operating Temperature 32 to 122 F 0 to 50 C Operating Humidity Range 0 to 85 Storage Temperature 14 to 122 F 10 to 50 C Dimensions 7 1 x 3 5 x 1 9 180 x 90 x 47mm Weight 8 2oz 500g REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com ...

Page 7: ...OFF Button 3 C F Selection Button 4 RJ ON Button T C Reference Junction Compensation Button 5 Function Selection Output Button 6 Range Selection Button 7 Zero Reset Button 8 Output Value Setting Button 9 Output Digit Selection Button 10 Output Terminals REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com ...

Page 8: ...ts reference cold junction compensation 7 Output State Indicator Operating Instructions Power ON OFF Press the POWER button to turn the instrument on Press and hold the POWER button to turn the instrument off When the instrument is on it will begin an internal self diagnosis at which time the full screen will be displayed When complete the instrument is ready to take measurements continued REED In...

Page 9: ...setting 7 Press and hold the POWER button to exit the maintenance state and turn the instrument off Output Function The output terminal of the instrument can produce DC voltages set by the user or can simulate resistance Do not apply any voltage to the output terminal during the operation If any improper voltage is applied to the output terminal it will cause damage to the internal circuit Simulat...

Page 10: ...t on the display 5 Press the and buttons to change the numerical value of each digit Numerical value cannot be changed beyond range maximum 6 Automatic compensation for reference junction temperature 7 Press the ZERO button and the output will be directly set to 0000 C R or S type 400 C B type or 0000 0 C other type 8 Press the C F button to select the unit C or F Note During the direct calibratio...

Page 11: ...ng output according to user s requirement Note An additional set of test leads are required for 3 wire or 4 wire systems 2 The display of the symbol OUTPUT denotes the instrument is in an output state 3 Press the FUNC button to select thermal resistance output or RTD output as indicated by Ω or C and Cu10 Copper 10 ohm RTD on the LCD display 4 While RTD function is selected press the RANG button t...

Page 12: ...e than 0 1μf the simulator will produce improper resistance Battery Replacement When the battery symbol appears on the display the battery needs to be replaced In order to replace the battery proceed with the following steps 1 Ensure that the meter is turned OFF and remove any test leads from the meter s terminals 2 Lift the tilt stand on the back of the unit to access the battery compartment door...

Page 13: ...ument Use on external parts only Product Warranty REED Instruments guarantees this instrument to be free of defects in material or workmanship for a period of one 1 year from date of shipment During the warranty period REED Instruments will repair or replace at no charge products or parts of a product that proves to be defective because of improper material or workmanship under normal use and main...

Page 14: ...EED distributor or REED Instruments Customer Service by phone at 1 877 849 2127 or by email at info reedinstruments com Please visit www REEDINSTRUMENTS com for the most up to date manuals datasheets product guides and software Product specifications subject to change without notice All rights reserved Any unauthorized copying or reproduction of this manual is strictly prohibited without prior wri...

Page 15: ...Test Measure with Confidence Check out our latest Products REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com ...

Page 16: ...REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com ...

Reviews: