8. The reading on the LCD screen is the value measured.
9. The reading can be corrected by removing the probe from the
layer and by pressing the plus or minus button.
10. Repeat step 7-10 until the result is correct.
Calibration Foils
This meter includes different foil sets for different ranges.
Please see the following table for reference.
Range (µm)
CM25
CM50
CM100 CM200 CM500 CM1000
0-200
•
•
•
•
0-500
•
•
•
•
0-1000
•
•
•
•
•
0-2000
•
•
•
•
•
Battery Replacement
1. When the low battery symbol appears in the display, it is time to
replace the batteries.
2. Turn the meter off, open the battery compartment, and remove
the batteries.
3. Install 4 new AA batteries verifying you are following the
correct polarity.
For service or information on this or any other REED product, contact
REED Instruments at [email protected].