INSTRUMENTS
Calibration Procedures
Calibrate the instrument before operation if the instrument was not in use
for a long time or it was operated in a bad environment
1. Make the following switch settings:
Display: dBA
Time weighting: FAST
Measurement mode: MAX MIN Mode function disabled
Level range: 50 to 100dB
2. Insert the microphone housing carefully into the insertion hole of
the calibrator.
3. Turn on the switch of calibrator and adjust the CAL potentiometer
of the unit. The level display will indicate the desired level.
4. This meter was calibrated before shipment. In order to ensure
the accuracy of the sound level meter for a long period of time it
should be calibrated once a year.
Setup TestLink RS232 Interface Software
System Requirements: Windows 98/NT 4.0/NT2000/XP/VISTA
Minimum Hardware Requirements: PC or NoteBook with Pentium 90MHz
or higher, 32 MB RAM. At least 5 MB hard drive space available to install
TestLink. The recommended resolution is 800 x 600.
Installing TestLink
1. Close all other applications before installing TestLink
2. Insert setup CD into the CD drive.
3. Choose the Start button on the Taskbar and then select Run
4. Type E:\SETUP and click OK, then it will copy SE322.exe
(executable file) as well as the help file to your hard drive
(default is c:\program files\TestLink\SE322)
5. For other operation instruction details, refer to the online help
while executing SE322