7
2.
Apply the measuring liquid to the sample well (minimum of 0.3ml).
Note:
For accurate measurements, it is recommended to completely cover
the sample well surface.
The temperature of the measuring liquid should be the same as the
ambient temperature. If the temperature is different, allow the liquid
to adjust to the ambient temperature before taking a measurement.
3.
When ready, close the protective cover and press the
START
button
to take a measurement.
4.
A series of dashes ( – – – ) will appear on the display followed by the
measured Brix value (%) and the temperature value of the liquid sample.
5.
Remove the liquid sample by wiping it off with a tissue properly
clean the sample well as highlighted in the
"Sample Well
Maintenance"
section.
Auto Power Off
To preserve battery life, the meter is programmed to turn off after
approximately 2 minutes of inactivity.
Sample Well Maintenance
Cleaning of the sample well should be performed immediately after each
sample reading. Never immerse the meter in any liquid. When the sample
well has been completely cleaned no residue should be present.
To properly clean the sample well use Isopropyl Alcohol followed by a
distilled water rinse and then thoroughly dry with the included cleaning
cloth, a residue free cloth or delicate task wipes.
Charging the Battery
1.
Connect the R9510 via the included cable to a USB port on your PC
or into a wall outlet using a USB Power Adapter (not included) to
charge the Li-ion battery.
Note:
The meter must remain powered off to properly charge via the
USB port on your PC.
2.
Charge the meter until the battery indicator appears full and remove
the charging cable when done.
REED Instruments
1-877-849-2127 | [email protected] | www.reedinstruments.com