D e t a i l e d P e r f o r m a n c e S p e c i f i c a t i o n s
Specification
iDVM 510 Accuracy
DC Voltage 600mV/6.0V/60V/1000V
+/- 0.5% +2 counts
DC Current 10A
+/- 0.8% + 2 counts
Capacitance 6.0nF/60nF/600nF/60µF/600µF/60mF/600mF
+/- 4% + 5 counts
AC Voltage 600mV/6.0V/60/1000V
+/-1.0% + 2 counts
AC Current 600µA/6000µA/60mA/600mA
+/- 1.2% + 2 counts
Frequency 0-60MHz
+/- 0.3% + 2 counts
DC Current 600µA/6000µA/60mA/600mA
+/- 0.8% + 2 counts
AC Current 600µA/6000µA/60mA/600mA
+/- 0.8% + 2 counts
AC Current 600µA/6000µA/60mA/600mA
+/- 3.0% + 2 counts
Continuity < 50 ohms
+/- 5 ohms
I n t e n d e d f o r U s e i n t h e F o l l o w i n g
E n v i r o n m e n t a l C o n d i t i o n s
Operating environment in which there is no pollution or only dry non-
conductive pollution - Pollution degree 2 (two) per IEC/EN61010
At altitudes up to 2,000 meters
At temperatures b5 to +35 degrees C
At maximum relative humidity of 80% for temperatures up to 31 degrees
C decreasing linearly to 50% relative humidity at 35 degree C
Up to 1000 Volts DC or AC RMS maximum voltage between any terminal
with transient over voltages of 8KV peak per IEC/EN61010