Redback Q 2003 Operating Manual Download Page 2

www.altronics.com.au

Redback® Proudly Made In Australia

2

* Specifications subject to change without notice. 

Distributed by Altronic Distributors Pty. Ltd. Perth. Western Australia.   Phone: 1300 780 999 Fax: 1300 790 999 

SYSTEM CHECK MODE

System  check  Mode  is  included  for  audible  testing  of 
larger speaker installations. It allows the Q 2003 to run at 
full output for a period of approximately 6 minutes so the 
installer can walk around the install to check sound is present 
on all speakers.
To activate this mode connect the test leads to the load to be
checked,  ensure  that  an  amplifier  is  not  connected.  Make 
sure the Q 2003 is OFF then press the TEST button and hold 
for  5  seconds.  “SYSTEM  CHECK”  will  display  on  the  LCD 
and a 1kHz tone will be put out the test leads.

SWITCHING OFF

To switch off the Q 2003 all you need to do is press the TEST 
button again. The LCD will display “OFF” for a short time 
then the unit will power down. The Q 2003 is also equipped 
with an auto power off function. The unit will automatically 
power  down  after  a  period  of  20  seconds  in  Single  test 
mode and 6 minutes in System check mode, saving precious 
battery life.

BATTERY INSTALLATION AND OR CHARGING

“LOW BATTERY” will display on the LCD when the battery 
requires replacing or if a rechargeable 9V Nicad/Nimh battery
(such as the Altronics S 4740C, not supplied) needs charging.
To  replace  or  install  the  9V  (type  216)  battery  required  to 
power  the  Impedance  Meter,  simply  slide  off  the  battery 
compartment on the back of the unit. Remove and unclip 
the old battery from the connector and replace it with a fresh 
battery. Slide the cover back until it locks into place.

To  recharge  an  installed  Nicad/Nimh  battery  connect  a 
12VDC source (car battery, cigarette lighter socket or plug 
pack)  >300mA  with  a  2.1mm  centre  positive  plug  to  the 
charging socket located on the top of the Q 2003.
“CHARGING” will display on the LCD along with the charge 
%. When the charge % reaches 100%,  the unit will stop 
charging and display “CHARGE FINISHED” on the LCD. The 
charger can then be unplugged. The inbuilt charging circuit 
is a slow charger and will take up to 12 hours to charge a 
completely flat battery.
NOTE:
1) The Q 2003 will not function as an Impedance meter when
charging.
2) The charger can be removed at anytime during charging 
though battery life will suffer.
3) The Q 2003 will still function when “LOW BATTERY” is 
being shown on the LCD but the measured results may be 
out by as much as 10%.

IMPORTANT GUIDELINES:

Don’t connect the meter to a line that has an amplifier 
attached to it. The impedance of the output transformer 
of an amplifier is extremely low and you could be easily 
be looking for a non-existing short.

Don’t connect the meter to an active line. You will 
damage the meter with the applied voltage.

Don’t continue to use a battery once the “LOW 
BATTERY” is being shown on the LCD. The readings 
you will get will be incorrect.

i

Wiring speakers in parallel for 70/100V line:  

Where several speakers are to

be used at one time, on one circuit, it becomes necessary to use speakers fitted
with line-matching transformers.  This is to overcome the effects of connecting
speakers in parallel and cable losses. The amplifier generally has an output
voltage of 100 volts (70 volts is typically used in North America, however
operation is similar). In this configuration the total wattage load on the amplifier
is derived from adding all the line transformer primary tap ratings together.  For
example, 70 one watt speakers will have a total speaker load of 70 watts.  Or
alternatively, it is conceivable to connect 100 one watt speakers to a 100 watt,
100 volt line amplifier.

Measuring 70/100V Line Speaker Impedance:  

To measure amplifier system

load, you must use an impedance meter in order to measure the ac resistance
of the connected speaker network. Impedance cannot be measured with a
standard multimeter, as this measures the dc resistance. Use the Altronics
Q 2001 or similar impedance meter.

About 70V & 100V Line Speaker Systems

Load        70V             100V

0.5W        9.4kΩ          20kΩ
0.66W      7.12kΩ        15kΩ
1W           4.7kΩ          10kΩ
1.25W      3.76kΩ        8kΩ
2W           2.35kΩ        5kΩ
2.5W        1.88kΩ        4kΩ
3W           1.56kΩ        3.3kΩ
5W           940Ω           2kΩ
7.5W        626Ω           1.3kΩ
10W         470Ω           1kΩ
15W         313Ω           666Ω
20W         235Ω           500Ω
30W         156Ω           333Ω
40W         117Ω           250Ω
60W         78Ω             166Ω
100W       47Ω             100Ω
125W       37Ω             80Ω
250W       19Ω             40Ω
500W       9.4Ω            20Ω

Reviews: