Error Flag E-CL
If error flag “E-CL” appears at power-up, a loss of calibration parameters
due to noise spikes has occurred. Entering code 77 twice in Factory Service
Operations Module (9-FS) erases the controller calibration values and
defaults the values to nominal settings. Reading errors of
±
10% may result. It
is recommended that the controller be fully re-calibrated. If using
thermocouple only, the RTD calibration need not be performed.
Note: The “E-CL” flag may be cleared by “stepping” through cold junction or
linear output calibration procedures without the need to change any
calibration values. A 10% reading error will still exist.
Calibration For T48
When re-calibration is required (generally every two years), this procedure
should be performed by qualified technicians using appropriate equipment.
Equipment source accuracy of 0.03% or better is required.
The procedure consists of: applying accurate mV signals, setting the
thermocouple cold junction temperature, and applying precision resistance,
among others. Allow a 30 minute warm-up period before starting this procedure.
Do not use thermocouple wire for the Millivolt or RTD ohms Calibration.
This procedure may be aborted by disconnecting power to the controller
before exiting the configuration mode. The existing calibration settings
remain in affect.
Note: After completing any of the calibration sequences, the controller defaults
the input sensor type to thermocouple type “J” (tc-j). Be sure to set input
sensor for proper type.
Note: The controller must be restored to normal display mode before any data
is stored.
Factory Service Operations - Calibration (9-FS)
Display
Parameter
Range
Description/Comments
CodE
Enter function code
48
Calibrate instrument
CAL
Millivolt Calibration
yes/no
Calibration required for both
RTD and TC input. If this
procedure is performed, the
cold junction temp or RTD
ohms calibration procedures in
turn must be completed.
CJC
Thermocouple cold
junction temperature
calibration
yes/no
Not required if only using RTD
input. This procedure can only
be performed AFTER an
accurate mV calibration.
rtd
RTD Resistance
Calibration
yes/no
Not required if only using TC
input. This procedure can only
be performed AFTER an
accurate mV calibration.
A1CL
Main Linear DC
Output Calibration
yes/no
Model number dependent.
Calibrate either 10 V or 20 mA
Range
A2CL
Second Linear DC
Output Calibration
yes/no
Model number dependent.
Calibrate either 10 V or 20 mA
Range
2CAL or
N2CL
Remote Setpoint or
Heater Current Input
Calibration
yes/no
Model number dependent
Millivolt Calibration (CAL)
Connect precision millivolt source with an accuracy of 0.03% to terminals
(-) #8 and (+) #9. Cold Junction or RTD Ohms calibration MUST be
performed after millivolt calibration.
Display
Action
Description/Comments
StP1
Apply 0.0 mV
Wait 10 seconds, press P.
StP2
Apply 14.0 mV
Wait 10 seconds, press P.
StP3
Apply 28.0 mV
Wait 10 seconds, press P.
StP4
Apply 42.0 mV
Wait 10 seconds, press P.
StP5
Apply 56.0 mV
Wait 10 seconds, press P.
-70-