background image

 

 

2  Operating instructions 

2.1  Button functions 

  RHC10, general purpose hand control. 

  RHC15, intended for patient lift systems. 

RHC10 and RCB15 standard features and design. RCH10 is available in 1-4 channel version 
They can be customized to fit unique needs. 

 

Figure 1. RHC10 2-channel version and RHC15 - control description  

2.2  Key lock function 

RHC10 has a Key lock function. To lock a channel press both buttons for that channel at the same time for 

3 seconds until the alert indicator LED is lit. Channel is locked. To unlock, do the same procedur. LED is 

switched off.  
Turning off the system with the on/off button and then start again, the red LED will light for a short while to 

indicate key lock is active.  
 

 

Summary of Contents for RHC10

Page 1: ...www reac se USER MANUAL RHC10 and RHC15 REAC HAND CONTROL...

Page 2: ...cation of advanced motion systems REAC s power solutions contain compact and strong electrical actuators lift and tilt systems control units and hand controls We know that our customers have different...

Page 3: ...TIONS 2 2 2 KEY LOCK FUNCTION 2 3 MAINTENANCE 3 3 1 SERVICE INTERVALS 3 3 2 CLEANING AND DISINFECTION 3 3 3 WASTE DISPOSAL 4 3 4 WARRANTY 4 3 5 REPAIRS 4 4 TROUBLE SHOOTING 5 5 TECHNICAL DATA 6 5 1 SY...

Page 4: ...data sheet RHC15 data sheet 1 2 Symbols The following symbols will be used in this document Failure to comply with these instructions may result in accidents involving personal injury Failing to follo...

Page 5: ...eds Figure 1 RHC10 2 channel version and RHC15 control description 2 2 Key lock function RHC10 has a Key lock function To lock a channel press both buttons for that channel at the same time for 3 seco...

Page 6: ...ables The cable must be checked for mechanical damage Cover The plastic housing must be checked for mechanical damage cracks Overlay The overlays must be checked for mechanical damage and firmly affix...

Page 7: ...There is a warranty on REAC products against manufacturing faults The warranty period begins on the purchasing date of the product and applies for 12 months Warranty exclusions REAC is entitled to den...

Page 8: ...act with control unit Nothing happens when pressing a button no LEDs are lit Make sure hand control is properly plugged in to control unit Low battery level in control system Nothing happens when pres...

Page 9: ...l units RCB10 RCB15 RCB11 RCB20 RCB12 RCB25 5 2 Basic characteristics RHC10 RHC15 No of actuator channels Up to 4 Up to 3 IP class IPX4 IPX4 Weight 0 2 kg 0 2 kg Color Black RAL9005 White RAL9016 Blac...

Page 10: ...ure 40 C to 70 C 40 C to 70 C Relative humidity 15 to 90 15 to 90 Atmospheric pressure 700 to 1060 hPa 700 to 1060 hPa Using the the products outside theire specified limits may lead to subsequent mal...

Page 11: ...sed on the label of REAC Control System products IPXX Ingress of particles first character and water second character as per EN60529 DC Voltage WEEE compliant CE label attached based on Low Voltage Di...

Page 12: ...Control System products Manufacturer manufacturing date Do not use if package is damaged Fragile handle with care Keep dry Temperature limits Relative humidity limits Atmospheric pressure limits Cons...

Page 13: ...10...

Page 14: ...always is up to date nor can we take responsibility for any misinterpretation of our written context Technical specification might change due to load and external circumstances REAC products shall be...