7
Cable Cores
Trim
Rubber Seal
Cable Shield Double Back
Over Plastic Sleeve
Plastic Sleeve
Cable
Gland Cap
Plastic Sleeve
Metal Gland
Nut
Wall of
Instrument Case
Fig. 4 Signal Cable Installation for Optimum EMC
1
2
Insert the end of the cable, plus the plastic sleeve into the metal outer shell of the
gland. The bore of the gland is a tight fit onto the cable shield, giving the required
ground contact.
3
Fit gland cap and tighten
2.4
TM202 on RDP transducers.
Some RDP transducers as standard have input impedance that is too low for use with the
S7TW. To correct this, a modification is available (TM202) for these transducers. The
modification code will be indicated on the transducer BUT if more than one modification is
applied to the transducer, the code may not be TM202. In this case, contact the factory.
List of RDP transducers requiring TM202.
D5/25,
D5/25H,
D5/25K,
D525HK
D5/2000,
ACT2000,
ACW2000 (with any armature configuration)
D5/4000,
ACT4000,
ACW4000 (with any armature configuration)
D5/15000, ACT15000, ACW15000 with any armature configuration)
D5/18500, ACT18500, ACW18500 (with any armature configuration)