Operation
28
Rev. A1 Jan 2015
3i Plus
10.
Push the <Next> soft key to move through each parameter.
11.
Push the <Done> soft key when finished. The display reverts to the start of the <Time/Date>
menu.
5.10.10.2
Alarm
The instrument has a programmable alarm to designate high or low readings, depending on the
thresholds entered. When the alarm level is reached, an alarm will sound and the display will flash
orange and white. To set either the high or low alarm:
1.
Push the <Menu> soft key until <Setup> shows as the right soft key function.
2.
Choose the <Alarm> soft key to access the <Alarm> menu.
3.
Push either the <Hi> or <Lo> soft key, depending on which alarm will be set.
4.
Push the <On> soft key to turn the alarm on.
5.
Push the <Off> soft key to turn the alarm off.
6.
Use the <Set> soft key to access the <Hi> or <Lo> Alarm Set menu.
7.
Use the down or up soft keys to change the alarm setting.
8.
Once the necessary settings have been entered, push the <Done> soft key.
5.10.10.3
Transmissivity
To change the <Transmissivity> value when measuring target temperature through glass windows:
1.
Push the <Menu> soft key until setup shows as the right soft key function.
2.
Choose <Transmissivity> to access the <Alarm> menu.
3.
Push the <On> soft key to turn the transmissivity on.
4.
Use the <Set> soft key to access the transmissivity set menu.
Transmissivity could be set between 1% and 100%.
5.
Push the <Off> soft key to turn <Transmissivity> off.
5.10.10.4
Background Temperature
This feature is useful when the target emissivity is below 1.0 and the background temperature is not
significantly lower than the target temperature. For instance, the higher temperature of a furnace wall
could lead to too-high temperatures being measured especially for lower emissivity targets. A built-in
ambient background temperature utility compensates for the impact of the reflected radiation in
accordance to the reflective behavior of the target. User can set this function as follows:
1.
Push the <Menu> soft key until setup shows as the right soft key function.
2.
Choose <Background Temperature> to access the <Alarm> menu.
3.
Push the <On> soft key to turn the <Background Temperature> on.
4.
Use the <Set> soft key to access the <Background Temperature Set> menu.
5.
Push the <Off> soft key to turn the <Background Temperature> off.
5.10.10.5
Bluetooth
Menu to turn Bluetooth on and off.
5.10.10.6
Language
To change the display language:
1.
From the main menu, push the <Menu> soft key until <Setup> shows as the left soft key
function.
2.
Push the <Setup> soft key.
3.
Use the down arrow soft key to move the indicator to <Language>.
4.
Push the <Enter> soft key.
5.
Push the <F1> soft key to complete the language selection or push the <Back> soft key to
return to the Setup menu.
1.888.475.5235