Raynger 3i Series Operator’s Manual
2-15
7. Pull the trigger further until you hear the “beep” indicating that the measurement
has been stored. Both the current and MAX temperatures are stored in memory.
Notes: The instrument has a 2-stage trigger. The first stage is activated by pulling the trigger
a small amount. The second stage (operational in LOG only) is activated when the
trigger is pulled all the way.
To protect the data from being accidentally overwritten (trigger accidently pulled),
press the RUN/LOG button to exit the LOG loop (the LOG icon is off).
8. Release the trigger.
LOG contains two modes: LOC (location) and MAX (as shown in Figure 2-7). In
LOG, the current temperature and emissivity setting along with either the location
number or MAX temperature can be displayed. Note the following:
• Press the MODE button to change between LOC and MAX modes.
• Press the
▲
or
▼
buttons to change the location number or the emissivity set-
ting.
• Press the backlight button if a brighter display is needed.
Notes: If you change the emissivity after storing data, both the target temperature and MAX
temperature will reset to zero. The instrument will “beep” to indicate this.
IMPORTANT
When you simultaneously press the MODE and ACTIVATE buttons
while in the LOG loop, all previously stored data is cleared. The
instrument also resets to its LOG loop factory default settings. The
instrument will “beep” after reinstalling the default settings. Section
3.6 lists default setting for each model as well as non-model-specific
default settings and range values.
1.888.475.5235
Summary of Contents for RAYNGER 3i LTDL2
Page 2: ...1 888 475 5235 info Raytek Direct com...
Page 4: ...1 888 475 5235 info Raytek Direct com...
Page 8: ...Table of Contents Raynger 3i Series Operator s Manual 1 888 475 5235 info Raytek Direct com...
Page 12: ...1 4 Raynger 3i Series Operator s Manual 1 888 475 5235 info Raytek Direct com...
Page 46: ...2 34 Raynger 3i Series Operator s Manual 1 888 475 5235 info Raytek Direct com...
Page 58: ...A 4 Raynger 3i Series Operator s Manual 1 888 475 5235 info Raytek Direct com...
Page 66: ...C 4 Raynger 3i Series Operator s Manual 1 888 475 5235 info Raytek Direct com...
Page 68: ...D 2 Raynger 3i Series Operator s Manual 1 888 475 5235 info Raytek Direct com...
Page 70: ...E 2 Raynger 3i Series Operator s Manual 1 888 475 5235 info Raytek Direct com...
Page 72: ...F 2 Raynger 3i Series Operator s Manual 1 888 475 5235 info Raytek Direct com...