34
Table 11 Instructions and solutionsfor alarms of laser
Alarm name
Alarm instructions and solutions
System Timer
Alarm
Instruction:
The internal clock of laser is abnormal.
Solution:
If it displays this alarm, please contact Raycus.
Laser On Button
Alarm
Instruction:
This alarm occurs when the “LASER” button on the front panel of the laser has
been pressed before the laser was powered on.
Solution:
Please power off the laser and pop upthe”LASER” button. Re-power the laser
and check if the alarm is released. If this alarm continues to occur, please
contact Raycus.
InterLock Alarm
Instruction:
Occurs when the InterLock is disconnected.
Solution:
Short the InterLock pins and restart the laser to try. If this error continues to
occur, please contact Raycus.
Current Driver
Alarm
Instruction:
Current Driver Alarm,this error occurs when the constant current driver board
inside the laser is abnormal.
Solution:
Restart the laser, If this error continues to occur, please contact Raycus
Laser Out Alarm
Instruction:
1. Laser Out Alarm immediately occurs when the laser is powered.
Solution:
The laser will autolock after two times of this error occurs successive, please
contact Raycus
2. The laser will get this error when it does not emit with correct settings.
Solution:
Close the shutter and pop up the “LASER” button on the front panel of the laser
to check if the red guide light is output normally when it isn’t emitting. If there
is no red guide light, please stop using the laser immediately and contact
Raycus. If the red guide light output normally, please restart the laser to try, if
the alarm continues to occur, please contact Raycus.
Laser Power
Alarm
Instruction:
The alarm is generated when the emission of the laser cannot reach the set
value. Power error occurs only when the laser is emitting.
Solution: