Instructions for Use: RAY® Osmotex Active Sterilising Facemask®
Version 2.1 16 April 2021
Connecting the electronic unit
There are two push buttons on the electronic unit to connect it to the mask. Use them to attach the
electronic unit to the mask.
Charging the electronic unit
Connect the charging cable to the electronic unit. Use a standard USB charger (protection class II,
CE/UL 5V/ minimum 30 mA).
Putting the mask into operation
RAY® Osmotex Active Sterilising Facemask® before each daily
use
Step1: charge the electronic unit SC-10. With the electroncis unit SC-10 fully charged, the mask can be
used for 10 hours.
Step2: before usage the mask should be wetted at the outside, by rinsing in luke warm water under the
tap or rinsing with a thoroughly wetted cloth after removing the electronic unit.
The wetting is needed in order to for the immediate onset of the electrolytic procees and
corresponding anti viral effect. If the mask is donned without pre-wetting, it will take about 30 half an
hour before the desinfection process starts due to moisture from exrpiration.
Please avoid direct contact between the electronic unit and moisture during when attaching it to
the mask. In case it gets wet, it will automatically turn off, and can be used again (by switching on in the
usual way) only after complete drying.
Step3: put the mask on with or without the electronic unit SC-10, holding it against your cheek in order
to cover nose and mouth. Adjust the shape of the nose clip with your hands. Fixate the mask by applying
the elastic bands / earloops behind your ears. Wear the mask comfortably by adjusting the lentght of
the elastic bands by means of the stoppers. Assure that the mask sits firmly.
Switching the electronic unit SC-10 on and off
Please use only the supplied electronic unit for operating
the mask! Please charge the electronic unit completely
before using it for the first time (see section Charging of
the electronic unit).
When the electronic unit is fully charged, press the button to the
right of the two LEDs for 1 second (see illustration).
The right diode lights up permanently green. The mask is now
supplied with permanent power and is activated. The fully charged
electronic unit supplies sufficient current (operating voltage: 0.6 V)
for at least up to 10 hours.
To switch off the electronic unit, press the button for 1-2 seconds
until the green LED goes off. The electronic unit is switched off and
can be activated by pressing the button again.