6
It is interesting to measure the amount of lift the sludge blanket experiences just
after the skimmer passes under the tank bridge. A steady hand lowering the
sensor probe will allow the operator to track the settling of the blanket after the
skimmer passes under the bridge. With experience, you will come to expect a
certain rate of settling after the skimmer passes. Unexpected settling differences
may indicate a process problem.
Always be aware of the skimmer location prior to lowering the sensor probe into
the tank.
Check for uniformity of blanket height by walking the length of the bridge
while maintaining the sensor probe just above the identified blanket.
Please feel free to share with Raven any other creative ways that you have
found to use this interactive tool in your process. Just logon to
Ravenep.com and leave us a note on the Contact page –
[email protected].
MAINTENANCE
Replace all 6 batteries as a group with new AA alkaline batteries of the
same type and brand.
Switch
SID
power switch-off when not in use to conserve batteries
Store
SID
for long periods with the batteries removed
Store
SID
in cool dry place and out of direct sunlight
Clean probe sensor eyes with mild detergent
TROUBLESHOOTING
NO POWER
– remove faceplate and verify connections between
battery cradle and printed circuit board.
NO POWER
– replace all six batteries with new of same brand
ALL RED LIGHTS ILLUMINATED
with no blockage of sensor probe
gap – indicates failure of signal cable. Likely point is near sensor
probe. Bend and wiggle signal cable along five foot section of cable