Figure 5-4
Wipe most-touched areas
7. Immediately insert the trap into the desorber.
8. Press Trig, the green button on the navigation pad, or the green button on the device handle.
The device automatically analyzes the sample.
•
If an alarm is detected, a message displays in the status bar.
Discard the
sample trap immediately, and follow your organization’s alarm resolution protocol.
•
If an alarm is not detected, the
No Alarm - Ready
message displays.
Remove the sample trap from the desorber.
By Hand Using a Sample Trap in a Sample Wand
The sample wand is used to sample in hard-to-reach areas and to protect the safety of the operator. The sample
trap is inserted into the clip portion of the sample wand. Proper installation ensures that the sample trap stays in
place.
1. Wearing clean cotton gloves, remove a sample trap from the package, or use an existing clean sample
trap that has not exceeded its sampling events.
2. If you obtained a sample trap from the package, close the cover immediately to avoid contamination.
3. Press the button on the sample wand, and insert the narrow end of the trap into the slot.
Avoid touching the sampling area of the sample trap when inserting it into or removing it from the sample
wand.
92
Rapiscan Systems Proprietary and Confidential
CHAPTER 5 │ User's Guide
Summary of Contents for Mobile Trace
Page 1: ...MobileTrace User s Guide For P0007027 MA001133 Revision ...
Page 2: ......
Page 31: ...This page intentionally left blank to ensure new chapters start on right odd number pages ...
Page 60: ...CHAPTER 4 Menu Functions Main Screen 44 Menu Functions 45 ...
Page 164: ...CHAPTER 9 Parts and Accessories Device and Replacement Part Numbers 148 ...
Page 167: ...This page intentionally left blank to ensure new chapters start on right odd number pages ...
Page 178: ...For P0007027 MA001133 Revision C ...