![Rankin Basics MCT45 Operator'S Manual Download Page 8](http://html.mh-extra.com/html/rankin/basics-mct45/basics-mct45_operators-manual_786398008.webp)
8
4. UNPACKING, TRANSPORTATION, AND STORAGE
4.1 Unpacking
•
Open the crate, take out all accessories and user manual.
Note: The transport crate and included
retaining elements should be kept in case a return shipment is necessary later.
•
Lift the instrument by holding it by the base plate and under the instrument on the
back (marked by blue circles in the pictures below) and lift it out of the formed cushion.
Do not hold the main and small handwheels and handles to lift the instrument!
4.2 Transportation
To guarantee trouble-free function of the instrument after transportation, please follow the instructions below
for transportation:
•
Shutting down the instrument for transportation: turn off the instrument, unplug the unit.
•
Remove all section waste debris and movable parts from the instrument.
•
Clean and disinfect all accessories according to the respective applicable lab regulations and transport
them in dry condition.
•
Unscrew the handwheel handle for transportation if necessary.
Note
Use original packaging materials for transportation of the instrument. If the original packaging is no
longer available, please contact customer service via
4.3 Storage
Storage conditions:
Storage environment: 0
℃
~
40
℃
Relative humidity should be lower than 80%
Atmospheric pressure 860hpa
~
1060hpa
Clean room without corrosive gas and well ventilated