Data management
R&S
®
ZNL/ZNLE
96
User Manual 1178.5966.02 ─ 19
Restrictions when recalling measurement settings
When recalling a saved configuration file, the following restrictions apply:
●
The R&S
ZNL/ZNLE must support the frequency range defined in the configuration
file.
●
Configuration files created on a R&S
ZNL/ZNLE with certain options in use do not
work on an R&S
ZNL/ZNLE without these options.
●
Files created with newer firmware versions may not work with a previous version.
●
Files created on an instrument other than the R&S
ZNL/ZNLE do not work on the
R&S
ZNL/ZNLE.
........................................................................................... 96
Configurable storage and recall
.............................................................................. 98
How to save and load instrument settings
............................................................ 103
6.2.1
Quick save/quick recall
The "Quick Save" and "Quick Recall" functions allow you to store instrument settings or
channel setups very easily and quickly in one step. Up to ten different sets of settings
can be stored to or recalled from "save sets". Each save set is identified by its storage
date and type (instrument or specific "Channel") in the display. The save sets are
stored in the
C:\Users\Public\Documents\Rohde-Schwarz\ZNL\QuickSave
directory, in files named
QuickSave1.dfl
to
QuickSave10.dfl
. Only the current
measurement settings are stored, not any additional data such as traces, limit line or
transducer files (see
Chapter 6.2.2.1, "Stored data types"
Source calibration files for an optional internal generator, if available, are included.
During recall, save sets of type "Instrument" replace the settings of the entire instru-
ment. All other save sets start a new channel setup with the stored settings.
If a channel setup with the same name as the "Channel" to be restored is already
active, the name for the new channel setup is extended by a consecutive number:
6.2.1.1
Quick save / quick recall settings
Access
: "Save"/ "Open" icon in the toolbar > "Quick Save" / "Quick Recall"
Both dialog boxes are very similar and closely related.
Storing and recalling instrument settings and measurement data