Command reference
R&S
®
ZNL/ZNLE
885
User Manual 1178.5966.02 ─ 19
Example:
CORR:CKIT:FOP 'N 50 Ohm','New Kit 1', '',
0,4000000000,0,0,50,99,-2.3,0.22,0,0,0,0,0
Define the properties of the Open (f) standard for a N 50 Ω con-
nector type in cal kit "New Kit 1".
CORR:CKIT:FOP? 'N 50 Ohm'
Query the properties of the Open (f) standard for a N 50 Ω con-
nector type in the active cal kit.
CORR:CKIT:FOP? 'N 50 Ohm','New Kit 1'
Query the properties of the Open (f) standard for a N 50 Ω con-
nector type in cal kit "New Kit 1".
Manual operation:
See
"One Port Standards / Two Port Standards"
Table 11-14: Set command parameters
One port standards
Two port standards
Open
(Offset)
Short
Match
Reflect
Sliding
match
Symmetric
network
Through,
Line
Attenuation
<Con-
nType>
mandatory
<CalKit-
Name>
mandatory (and must not be an empty string)
<Standar-
dLabel>
mandatory (but can be an empty string)
<MinFreq>,
<MaxFreq>
mandatory
<Delay-
Param>,
<Loss>,
<Z0>
mandatory
mandatory
mandatory
mandatory
not used
mandatory
not used
<C0>,
<C1>,
<C2>, <C3>
mandatory
mandatory
*
mandatory
*
mandatory
**
mandatory
**
not used
<L0>, <L1>,
<L2>, <L3>
mandatory
*
mandatory
mandatory
*
mandatory
**
mandatory
**
OPEN |
SHORt |
MATCh |
<Resist-
ance>
optional
***
optional
***
optional
***
mandatory:
OPEN |
SHORt |
<Resist-
ance>
mandatory:
OPEN |
SHORt |
<Resist-
ance>
<Port1>[,
<Port2>]
optional: <Port 1>
optional: <Port 1>,<Port 2>
*
values are ignored during calibration
**
if OPEN is selected the residual inductance (L0,...,L3) are during calibration; if SHORt is selected the fringing capacitance
(C0,...,C3) are during calibration
***
must be provided if the definition is restricted to a single port. Values are ignored during calibration
VNA command reference