Command reference
R&S
®
ZNL/ZNLE
825
User Manual 1178.5966.02 ─ 19
Parameters:
<TraceName>
DALL
–
all data traces
MALL
–
all memory traces
<string>
–
single trace identified by its trace name (string param-
eter), e.g. 'Trc4'.
<Boolean>
ON | OFF
–
display or hide traces.
Example:
*RST; :DISP:TRAC:SHOW? 'Trc1'
Reset the analyzer, creating the default trace 'Trc1'. The trace is
displayed; the query returns 1.
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named Ch4Tr1 to measure the
input reflection coefficient S
11
.
DISP:WIND2:STAT ON; :DISP:WIND2:TRAC:FEED
'CH4TR1'
Create diagram area no. 2 and display the generated trace in
the diagram area.
DISP:TRAC:SHOW DALL, OFF
Hide both traces in both diagrams.
DISP:TRAC:SHOW? DALL
Query whether all data traces are displayed. The response 0
means that at least one trace is hidden.
Manual operation:
See
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:X:OFFSet
<StimulusOffset>
Shifts the trace
<WndTr>
in horizontal direction, leaving the positions of all markers
unchanged.
Suffix:
<Wnd>
.
Number of an existing diagram area (defined by means of
ON
).
<WndTr>
Existing trace number, assigned by means of
WINDow<Wnd>]:TRACe<WndTr>:FEED
Parameters:
<StimulusOffset>
Stimulus offset value.
The range and unit depends on the sweep type.
Default unit: NN
Example:
*RST; :DISP:WIND:TRAC:X:OFFS 1MHZ; :DISP:WIND:
TRAC:Y:OFFS 10
Create the default trace and shift it horizontally by 1 MHz, verti-
cally by 10 dB.
Manual operation:
See
VNA command reference