Command reference
R&S
®
ZNL/ZNLE
614
User Manual 1178.5966.02 ─ 19
SYSTem:COMMunicate:GPIB[:SELF]:RTERminator
<Terminator>
This command selects the GPIB receive terminator.
Output of binary data from the instrument to the control computer does not require
such a terminator change.
Parameters:
<Terminator>
LFEOI | EOI
LFEOI
According to the standard, the terminator in ASCII is <LF>
and/or <EOI>.
EOI
For binary data transfers (e.g. trace data) from the control com-
puter to the instrument, the binary code used for <LF> might be
included in the binary data block, and therefore should not be
interpreted as a terminator in this particular case. This can be
avoided by using only the receive terminator
EOI
.
*RST:
LFEOI
Example:
SYST:COMM:GPIB:RTER EOI
Manual operation:
See
SYSTem:COMMunicate:INTernal:REMote
<State>
The instrument switches between manual and remote operation.
Note
: If the local lockout function (
LLO
) has been activated in the
remote control mode, manual operation is no longer available until
GTL
(or
) is executed.
For details, see
Chapter 8.4.6, "Returning to manual mode ("local")"
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
The instrument switches from remote to manual operation. You
can operate the instrument locally.
ON | 1
The instrument switches from manual to remote operation.
*RST:
0
Example:
SYST:COMM:INT:REM OFF
The instrument switches from remote to manual operation (cor-
responds to @LOC or selecting the "Local" softkey).
Manual operation:
See
Common instrument commands