Command reference
R&S
®
ZNL/ZNLE
606
User Manual 1178.5966.02 ─ 19
Example:
//Schedule a self-alignment every Sunday at 2 AM after a
warmup period, then shut down instrument.
CAL:DUE:WARM
CAL:DUE:SCH ON
CAL:DUE:DAYS SUND
CAL:DUE:TIME '2:00'
CAL:DUE:SHUT
CALibration:RESult?
This command returns the results collected during calibration.
Return values:
<CalibrationData>
String containing the calibration data.
Example:
CAL:RES?
would return, e.g.
Total Calibration Status:
PASSED, Date (dd/mm/yyyy): 12/07/2004,
Time: 16:24:54,Runtime: 00.06
Usage:
Query only
DIAGnostic:SERVice:INPut:PULSed:CFRequency
<Frequency>
This command defines the frequency of the calibration signal.
Before you can use the command, you have to feed in a calibration signal with
DIAGnostic:SERVice:INPut[:SELect]
Manual operation:
See
DIAGnostic:SERVice:INPut:PULSed:MCFRequency
<Frequency>
This command sets the calibration frequency for frequencies greater than 7 GHz. This
command only takes effect if a microwave calibration signal is selected for input
(
DIAGnostic:SERVice:INPut[:SELect]
DIAGnostic:SERVice:INPut[:SELect]
<Signal>
This command activates or deactivates the use of an internal calibration signal as input
for the R&S
ZNL/ZNLE.
Parameters:
<Signal>
CALibration
Uses the calibration signal as RF input.
MCALibration
Uses the calibration signal for the microwave range as RF input.
RF
Uses the signal from the RF input.
Common instrument commands