VNA GUI reference
R&S
®
ZNL/ZNLE
325
User Manual 1178.5966.02 ─ 19
Access:
[Sweep] hardkey
Background information
Refer to the following sections:
●
Chapter 9.1.4, "Sweep control"
●
Chapter 9.1.4.2, "Stimulus and sweep types"
10.5.1
Sweep Params tab
Allows you to define the scope and timing of the measurement in the active channel.
Segmented sweeps
In "Segmented" linear frequency sweeps, the sweep parameters can be set independ-
ently for each segment. See
Chapter 10.5.2.2, "Define Segments dialog"
System error correction
In general, the system error correction is no longer valid after a change of the sweep
parameters. The status of the calibration is shown in the trace list. If the number of
points is changed, the analyzer interpolates the correction data. The calibration label
"Cal Int" is displayed. See also
Chapter 9.5.4, "Calibration state labels"
Number of Points
Sets the total number of measurement points per sweep. The minimum number of
points is 1 (measurement at a single frequency/power/time value), the maximum is
100,001.
Together with the sweep range defined in the
of the "Stimulus" softtool,
this parameter defines the grid of sweep points. The step width between two consecu-
tive sweep points is constant on a linear scale (sweep types "Lin Freq", "Time" and
"CW Mode") or on a logarithmic scale (sweep types "Log Freq" and "Power").
Sweep Softtool