Instrument Functions
R&S
®
ZNH
82
User Manual 1334.5985.02 ─ 02
5.4
Presetting the R&S
ZNH
Before you prepare a measurement, it is recommended to preset the R&S
ZNH. During
a preset, the R&S
ZNH resets all settings to their default state. Restoring the default
configuration has the advantage that old settings do not affect measurements.
The default setup is specific to the operating mode.
► Press [PRESET] key.
The R&S
ZNH restores its default setup.
You can load or discard a previous calibration data during a preset.
You can also define your own default settings via a dataset. This dataset is then loaded
after pressing the [PRESET] key instead of the factory default.
1. Press [SETUP] key.
2. Select "User Preference" softkey.
3. In the "User Preferences" dialog box, select the "Preset Dataset" item.
The R&S
ZNH displays the file manager dialog box to select the dataset that con-
tains the settings you would like to have as the preset settings.
4. Select the dataset with the settings you want and press "Load" softkey.
5. In the "User Preferences" dialog box, select the "Preset Mode" item.
A drop-down menu opens to select the preset mode.
6. Select "User Defined" to load the dataset defined in the "Preset Dataset".
The R&S
ZNH now loads the settings of the dataset after you press [PRESET.]
7. Select "Discard Calibration Data" to define the usage of user calibration data after
preset.
Presetting the R&S
ZNH