Getting Started
R&S
®
ZNH
26
User Manual 1334.5985.02 ─ 02
Risk of instrument damage
To avoid damage to the tracking generator output, never apply currents greater than
600 mA or voltages greater than 20 V to the BNC connectors if the BNC connectors
are not configured as BIAS output ports.
If they are configured as BIAS ports, never apply currents greater than 600 mA or vol-
tages greater than 50 V.
Headphone jack
The 3.5 mm connector for headphones has an internal impedance of approximately 10
Ω.
USB type A connector
The USB port provides the USB interface to connect a memory stick and store data
sets or screenshots. It can also be used to control the operation of the external power
sensor.
Risk of electrical shock
To avoid electrical shock, the DC input voltage must never exceed the value specified
on the housing.
Risk of instrument damage
●
Make sure not to overload the RF input and keep within the maximum allowed sig-
nal levels. Refer to the datasheet for the maximum allowed signal levels.
●
To avoid damage to the coupling capacitor, input attenuator or the mixer, the DC
input voltage of 50 V must never be exceeded.
3.2.3
Left View
Instrument Tour