Remote Commands
R&S
®
ZNH
261
User Manual 1334.5985.02 ─ 02
Parameters:
<SearchLimit>
Sets the left marker search limit.
The unit in the spectrum analyzer mode depends on the mea-
surement, e.g. Hz for measurements in the frequency domain
and seconds for measurements in the time domain.
Range:
Maximum span.
*RST:
– (is set to the left diagram border when switching
on search limits)
Default unit: GHz
Example:
See
CALCulate<1...2>:MARKer<1...6>:X:
CALCulate<1...2>:MARKer<1...6>:X:SLIMits<1...2>:RIGHt
<SearchLimit>
This command defines the right limit of the marker search range.
To use the command, you first have to turn on search limits with
CALCulate<1...2>:MARKer<1...6>:X:SLIMits<1...2>[:STATe]
The numeric suffix at CALCulate is irrelevant for this command.
The numeric suffix at MARKer selects the marker. The numeric suffix at SLIMits selects
the limit line.
Parameters:
<SearchLimit>
Sets the right marker search limit.
The unit depends on the measurement, e.g. Hz for measure-
ments in the frequency domain and seconds for measurements
in the time domain.
Range:
Maximum span.
*RST:
– (is set to the right diagram border when switching
on search limits)
Default unit: GHz
Example:
CALC:MARK:X:SLIM ON
CALC:MARK:X:SLIM:LEFT 10MHz
CALC:MARK:X:SLIM:RIGH 100MHz
Turns on search limits and defines a search range from 10 MHz
to 100 MHz.
CALCulate<1...2>:MARKer<1...6>:Y?
This command queries the absolute vertical position of a marker.
If necessary, the corresponding marker is activated first.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end between activating the delta marker and reading out the result. This is only
possible in single sweep mode.
The unit of the return value depends on
UNIT<1...2>:POWer
.
Remote Control - Commands