Command Reference
R&S
®
ZNC
536
User Manual 1173.9557.02 ─ 13
Parameters:
<Boolean>
ON | OFF - enables or disables the automatic calculation of the
sweep average over the specified number of sweeps (
*RST:
ON
Example:
See
Manual operation:
See
6.3.14.2
[SENSe:]BANDwidth...
The
[SENSe:]BANDwidth...
commands set the bandwidth of the IF measurement
filter (resolution bandwidth).The forms
BANDwidth
and
BWIDth
are equivalent.
............................................................................536
.................................................................................536
[SENSe<Ch>:]BANDwidth[:RESolution]
<ResBandw>
[SENSe<Ch>:]BWIDth[:RESolution]
<ResBandw>
Defines the resolution bandwidth of the analyzer (Meas. Bandwidth). Values between 1
Hz and 100 kHz can be set.
Bandwidths can be set in 1
–
1.5
–
2
–
3
–
5
–
7 steps. The analyzer rounds up any entered
value between these steps and rounds down values exceeding the maximum bandwidth.
Suffix:
<Ch>
.
Channel number
Parameters:
<ResBandw>
Resolution bandwidth
Range:
See above
Increment: 1-1.5-2-3-5-7 steps
*RST:
10 kHz
Default unit: Hz
Example:
BAND 1.1
Set a resolution bandwidth of approx. 1.1 Hz for channel 1.
BAND?
The analyzer returns the rounded bandwidth of 1.5 Hz.
Manual operation:
See
6.3.14.3
[SENSe:]CORRection:CDATa...
The
[SENSe:]CORRection:CDATa...
commands read or write system error correc-
tion data.
SCPI Command Reference