Operating external frontends
R&S
®
SMW-K553
14
User Manual 1179.4240.02 ─ 02
To connect an R&S
FE50DTR
The R&S
FE50DTR and R&S
SMW are switched on and connected to power. At the
R&S
FE50DTR, you connect the IF input signal from the R&S
SMW, to the "IF In A" or
"IF In B" connector.
► To connect an R&S
FE50DTR with the R&S
SMW, follow the instructions described
and
provide an overview on connections and a test setup.
Table 3-2: R&S
SMW and R&S
FE50DTR connections
Signal
R&S
SMW
R&S
FE50DTR
Reference
Connector: "REF OUT"
Frequency: 1
GHz, 10
MHz
"Ref In"
LAN (Control)
"LAN"
"LAN"
IF In
"RF A"
"IF In B"
RF Out
-
"RF Out B"
DUT
RF shield box
R&S FSV/A
R&S SMx
RF
Ref Out
Ref Out
RF
LAN
IF In
Ref
IF Out
IF Out A
LAN
IF In B
RF A
RF B
Ref In
LAN
LAN
RF Out
LAN
RF In
R&S
FE50DTR
Figure 3-2: Test setup: R&S
SMx = R&S
SMW, R&S FSV/A, R&S
FE50DTR and RF shield box
For a more general test setup, see
3.2
Controlling external frontends
To configure the R&S
SMW for frontend control
The following procedure describes how to configure the R&S
SMW to establish a con-
trol connection between R&S
SMW and an external frontend.
The R&S
SMW is connected to the R&S
FE44S, see
on page 12. Configuration with other connected external frontends is analogous.
1. Select "Taskbar > System Config > System Configuration > External RF und I/Q".
2. Select "RF A > External Instrument > Config".
The "External Instrument Configuration" dialog opens.
Controlling external frontends