Desired Crest Factor
Enters the desired crest factor.
This is only possible when the optimization "Target Crest" has been selected.
Remote command:
Table Setup Assistant
The "Table Setup Assistant" section can be used to set a selectable carrier range. The
carrier table can be edited in the "Carrier Table" dialog.
Carrier Start
Defines the start index of the carrier range to which the following settings are intended
to apply.
Remote command:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:STARt
Carrier Stop
Defines the stop index of the carrier range to which the following settings are intended
to apply.
Remote command:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:STOP
Carrier State
Switches the carriers in the carrier range on/off.
Remote command:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:STATe
Power Start
Sets the power of the starting carrier.
Remote command:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:POWer[:STARt]
Power Step
Sets the width of the step with which the power will be changed from carrier to carrier.
The carrier power that is set with
Power + n* Power Step
must be within the valid
value range -80 dB to 0 dB.
Remote command:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:POWer:STEP
Phase Start
Sets the phase of the starting carrier. This setting is only available for "Optimize Crest
Factor Mode" = Off.
Remote command:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:PHASe[:STARt]
Baseband Signal - Baseband Block
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...