Instrument Function
R&S
®
SMJ100A
392
Operating Manual 1403.7458.32 ─ 14
Generate Signal HD (Sine Test Signal)
Generates a signal and saves it to a file.
The File Select window opens automatically and the signal can be stored as a wave-
form file.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:CREate:NAMed
Generate Signal RAM (Sine Test Signal)
Generates a signal and uses it as output straight away.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:CREate
Rect Test Signal
To access the "Sine Test Signals" dialog, in the "ARB MOD" dialog set the parameter
"Create Test Signal" to Rect and press "Create Test Signal".
The dialog can be used to configure a rectangular test signal. A rectangular test signal
with a duty factor of 0.5 is created. Amplitude and offset are adjustable. Both paths, I
and Q, get the same signal.
Frequency (Rect Test Signal)
Enters the frequency of the test signal.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:FREQuency
Samples per Period (Rect Test Signal)
Enters the number of sample values required for the rectangular signal per period.
The resulting clock rate must not exceed the maximum ARB clock rate (see data
sheet). The number of sample values is automatically restricted by reference to the set
frequency.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:SAMPles
Amplitude (Rect Test Signal)
Enters the digital amplitude of the rectangular wave.
The abbreviation FS means full scale.
Baseband Signal - Baseband Block
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...