Instrument Function
R&S
®
SMJ100A
301
Operating Manual 1403.7458.32 ─ 14
Note:
The peak envelope power (PEP) specified in the header corresponds to the PEP
value of the carrier.
In the mode "Additive Noise" the "Display Output Results For" (analog or digital) has to
be set before.
Remote command:
[:SOURce<hw>]:AWGN:POWer:SUM:PEP?
5.7.3 Output of the Baseband Signal - I/Q Out
The R&S
SMJ provides various outputs of the baseband signal.
The digitally modulated signal can either be output directly or converted to an analog
I/Q baseband output signal. The
analog baseband signal
can be output single-ended
or differential (non-inverted).
The single ended or differential output requires option R&S
SMJ-B16, Differential Out-
put.
The analog signal is output at the I/Q output connectors I OUT/I OUT BAR and Q OUT/
Q OUT BAR on the rear panel of the instrument.
The output of the
digital baseband signal
requires the option R&S
SMJ-B18, Digital
I/Q Out. The signal is output at the standardized (LVDS) digital I/Q interface on the rear
panel.
5.7.3.1
I/Q Out Settings Menu
The settings dialogs for the analog and digital signal output are available in the block
diagram in the "AWGN/IMP" function block as well as in the menu with the same name
which is opened using the MENU key.
Impairment of Digital I/Q Signal and Noise Generator - AWGN/IMP Block
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...