Remote Control Commands
R&S
®
SMJ100A
579
Operating Manual 1403.7458.32 ─ 14
[:SOURce<hw>]:AWGN:POWer:SUM?
In Additive Noise and CW Interferer (
SOUR:AWGN:MODE ADD|CW
) modes, queries the
overall power of the noise signal plus useful signal.
Tip:
Use the command
[:SOURce<hw>]:AWGN:DISP:ORESults
log or digital signal path before.
Return values:
<Sum>
float
Range:
-145 to 20
Increment: 0.01
*RST:
0
Example:
SOURce:AWGN:POWer:SUM?
queries the overall power of the noise signal plus useful signal.
Usage:
Query only
Manual operation:
See
"C Noise Power / Noise Power (System
Bandwidth) / C Interferer Power / Interferer
Power"
[:SOURce<hw>]:AWGN:POWer:SUM:PEP?
In Additive Noise and CW Interferer (
SOUR:AWGN:MODE ADD|CW
) modes, queries the
peak envelope power of the overall signal comprised of noise signal plus useful signal.
Tip:
Use the command
[:SOURce<hw>]:AWGN:DISP:ORESults
log or digital signal path before.
Return values:
<Pep>
float
Range:
-145 to 20
Increment: 0.01
*RST:
0
Example:
SOURce:AWGN:POWer:SUM:PEP?
queries the peak envelope power of the overall signal.
Usage:
Query only
Manual operation:
See
"C Noise PEP / Noise PEP (Total Band-
width) / C Interferer PEP / Interferer PEP"
[:SOURce<hw>]:AWGN:STATe
<State>
Activates or deactivates the white noise (AWGN = Averaged White Gaussian Noise).
Use the command
to define the mode the AWGN gen-
erator is working in.
SOURce Subsystem
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...