Instrument Function
R&S
®
SMJ100A
141
Operating Manual 1403.7458.32 ─ 14
Check that the key in the dialog turns red.
Note:
Pressing the same key again has no further effect, with the exception of the
ESC key, see
Press the ESC key a third time
.
5. Continue with the next key on the front panel and repeat
to
until all
keys are tested.
The test is completed, when each key is verified successfully, confirmed by a "Test
passed" message.
Select "OK" to exit the test.
► Press the ESC key a third time.
Exits the "Check Front Panel" dialog, even if you have not yet checked all the keys.
Expected responses:
●
Pressing a key once (green), pressing twice (red)
●
Pressing the ESC key a third time exits the dialog.
If you detect a malfunction, for example, you press the front panel key the first time,
and the color of the button in the dialog turns red (instead of green), the front panel key
may be stuck. In this case, contact the Rohde
&
Schwarz Customer Support Center for
.
5.2.3.10
Global Trigger/Clock/External Input Settings
The "Global Trigger/Clock/External Inputs" menu can be opened using the SETUP or
MENU key under the "Setup" menu as well as in all clock and trigger menus of the
"Baseband" function block.
This menu is used to determine the physical characteristics of the input connectors for
trigger, clock and control signals.
The user inputs/outputs (BNC connector and pins of the AUX I/O connector) are
assigned in the dialog
A common trigger threshold and input impedance is effective for all trigger and control
signal inputs. The settings influence the digital modulations, the generation of wave-
forms or multicarrier signals, and all digital standards. Irrespective of this, a common
threshold and input impedance is effective for the serial and parallel data input. These
data sources are available for digital modulation ("Custom Digital Modulation").
The instrument trigger setting influences all sweeps and is effective in the List mode.
General Instrument Settings
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...