5.2.3.1
Internal Adjustments
The R&S
SMJ is extremely accurate due to the integrated adjustment procedures,
which you can execute directly on the instrument.
Self-calibration routines that do require additional (external) measurement equipment
are either described in the Service Manual of the instrument, or they require to be per-
formed by a Rohde
&
Schwarz service center.
Risk of invalid adjustment
In order to achieve correct adjustment of the instrument, make sure that the instrument
is warm before performing adjustments. The warm-up time is 30 minutes.
When to perform adjustments?
It is recommended to perform internal adjustments prior to any application that requires
a maximum of level and frequency accuracy, especially when a long period of time has
passed or if the ambient temperature of the instrument significantly differs from the one
of the last adjustment.
Additional Information to the Adjustments
During adjustments, a progress indicator shows the status of the process. If any error
occurs, the process aborts and an error message appears in the info line.
The extent of the adjustments depends on the installed options.
► To access the settings for internal adjusmtents, press the SETUP key and select
"System > Internal Adjustments".
General Instrument Settings
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...