Remote Control Commands
R&S
®
SMB100A
448
Operating Manual 1407.0806.32 ─ 21
Setting parameters:
<TimeMs>
integer
Wait time in ms
Range:
0 to 10000
*RST:
0
Example:
:SYSTem:WAIT 10000
// waits 10s before resetting the instrument
*RST
Usage:
Setting only
7.16 TEST Subsystem
The TEST system contains the commands for performing selftest routines, and for
direct adjustment of the hardware assemblies (
:TEST:DIRect
).
The self tests return a "0" if the test is performed successfully, otherwise a value other
than "0" is returned. None of the commands of this system have an
*RST
value.
Improper use can destroy the assembly
The respective hardware assembly responds directly to the
:TEST:DIRect
command;
any safety mechanisms are bypassed. The command is intended for servicing purpo-
ses and should be used only by the Rohde
&
Schwarz service personnel.
................................................................................................448
:TEST<hw>:ALL:STARt
Starts a self-test on all installed hardware options.
To query the result, use the command
Example:
See
Usage:
Event
Manual operation:
See
:TEST<hw>:ALL:RESult?
Queries the result of the performed self-test (command
TEST Subsystem