Preparing for Use
R&S
®
SMB100A
29
Operating Manual 1407.0806.32 ─ 21
2.3.5.1
Standby and Ready state
The ON/STANDBY key is located in the bottom left corner of the front panel.
Switching between standby and ready state
► Press the ON/STANDBY key briefly to switch the instrument from the standby to
ready state or vice versa.
In ready state, the left, green LED is on. The instrument is ready for operation. All
modules are power-supplied and the R&S
In standby state, the right, yellow LED is on. The standby power only supplies the
power switch circuits and the oven-controlled crystal oscillator to keep it at its oper-
ating temperature. In this state it is safe to switch off the AC power and disconnect
the instrument from the power supply.
2.3.5.2
Start Display and Booting
The instrument boots the operating system, starts the instrument firmware and auto-
matically performs a selftest. If the previous session was terminated regularly, the
instrument uses the last setup with the relevant instrument settings.
Once the startup procedure has been terminated, the block diagram opened in the pre-
vious session is displayed and the instrument is ready for operation.
Use the PRESET key to return the instrument to its definite reset/preset state, if the
current setup is not anymore relevant.
To customize the start settings, use the "File" dialog.
Rebooting the instrument
Press the STANDBY key for approx. 5 s. to reboot the instrument, if the software stops
unexpectedly.
2.3.5.3
Function Check
The signal generator automatically monitors the main functions when the instrument is
switched on and continuously during operation.
A detected fault is indicated by a "ERROR" message displayed in the info line together
with a brief error description. For in-depth identification of the error, press the INFO
key. In response, a description of the error(s) is displayed.
For more information, refer to
Chapter 9, "Status Information, Error Messages and
Additionaly to the automatic monitoring, the R&S
SMB offers the following capabilities
to ensure correct functioning:
●
Internal Adjustments
Putting into Operation